Tue Jan 4 18:26:59 2022, Anchal, Update, BHD, Tested 2" PR2 candidates transmission
|
Wed Jan 5 17:46:04 2022, Anchal, Update, BHD, Tested 2" PR2 candidates transmission
|
Mon Jan 10 18:20:45 2022, Anchal, Update, BHD, Tested 2" PR2 candidates transmission
|
Tue Jan 11 15:21:16 2022, Anchal, Update, BHD, PR2 transmission calculation  
|
Thu Jan 13 17:10:55 2022, Anchal, Update, BHD, PR2 transmission calculation  
|
Fri Jan 14 03:07:04 2022, Koji, Update, BHD, PR2 transmission calculation
|
Fri Jan 14 11:00:29 2022, Anchal, Update, BHD, PR2 transmission calculation
|
Fri Jan 14 13:46:25 2022, Anchal, Update, BHD, PR2 transmission calculation updated  
|
Wed Jan 19 16:22:48 2022, Anchal, Update, BHD, PR2 transmission calculation updated 
|
Thu Jan 20 01:48:02 2022, Koji, Update, BHD, PR2 transmission calculation updated
|
|
Message ID: 16543
Entry time: Wed Jan 5 17:46:04 2022
In reply to: 16541
Reply to this: 16566
|
Author: |
Anchal |
Type: |
Update |
Category: |
BHD |
Subject: |
Tested 2" PR2 candidates transmission |
|
|
I tested 2 more optics today, the old PR2 that we took out and another optic I found in QIL. Both these optics are also not good for our purpose.
|
Polarization |
Incident Power [mW] |
Transmitted Power [mW] |
Transmission [ppm] |
Existing PR2 |
p-pol |
910 |
0.004 |
4.4 |
V2-1698 & V2-1700 |
p-pol |
910 |
595 |
653846 |
I'll find thw Y1S optic and test that too. We should start looking for alternate solutions as well.
|