Message ID: 16312     Entry time: Thu Sep 2 21:21:14 2021     In reply to: 16307
Author: Koji 
Type: Summary 
Category: Computers 
Subject: Vacuum recovery 2 

Attachment 1:
We are pumping the main volume with TP2. Once P1a reached the pressure ~2.2mtorr, we could open the PSL shutter. The TP2 voltage went up once but came down to ~20V. It's close to nominal now.
We wondered if we should use TP3 or not. I checked the vacuum pressure trends and found that the annulus pressures were going up. So we decided to open the annulus valves.

Attachment 2:
The current vacuum status is as shown in the MEDM screenshot.

There is no trend data of the valve status (sad)