40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue May 25 17:42:11 2021, Anchal, Paco, Summary, ALS, ALS Single Arm Noise Budget ALS_IR_b.svgALS_Single_Arm_IR.pdf
    Reply  Thu May 27 11:03:15 2021, Anchal, Paco, Summary, ALS, ALS Single Arm Noise Budget ALS_Single_X_Arm_IR.pdfALS_OOL_with_Ref.pdf
       Reply  Fri May 28 17:32:48 2021, Anchal, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat SingleArmActCalwithIRALSBeat.pdfstateSpaceModel.zip
          Reply  Tue Jun 1 16:55:32 2021, Anchal, Paco, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat SingleArmActCalwithIRALSBeat-1306624785.pdf
             Reply  Tue Jun 8 11:40:53 2021, Anchal, Paco, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat ITMX_Cal_Noise_Spectrum_1307143423.pdf
                Reply  Wed Jun 9 18:29:13 2021, Anchal, Paco, Summary, ALS, Check for saturation in ITMX SOS Driver 
                Reply  Fri Jul 9 15:39:08 2021, Anchal, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat [Correction] ITMX_calibration_With_ALS_Beat.pdf
Message ID: 16196     Entry time: Wed Jun 9 18:29:13 2021     In reply to: 16192
Author: Anchal, Paco 
Type: Summary 
Category: ALS 
Subject: Check for saturation in ITMX SOS Driver 

We did a quick check to make sure there is no saturation in the C1:SUS-ITMX_LSC_EXC analog path. For this, we looked at the SOS driver output monitors from the 1X4 chassis near MC2 on a scope. We found that even with 600 x 10 = 6000 counts of our 619 Hz excitation these outputs in particular are not saturating (highest mon signal was UL coil with 5.2 Vpp). In comparison, the calibration trials we have done before had 600 counts of amplitude, so we can safely increase our oscillator strength by that much yes


Things that remain to be investigated -->

  • What is the actual saturation level?
  • Two-tone intermodulation?