40m
QIL
Cryo_Lab
CTN
SUS_Lab
CAML
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Tue May 25 17:42:11 2021, Anchal, Paco, Summary, ALS, ALS Single Arm Noise Budget 
|
Thu May 27 11:03:15 2021, Anchal, Paco, Summary, ALS, ALS Single Arm Noise Budget 
|
Fri May 28 17:32:48 2021, Anchal, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat 
|
Tue Jun 1 16:55:32 2021, Anchal, Paco, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat
|
Tue Jun 8 11:40:53 2021, Anchal, Paco, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat
|
Wed Jun 9 18:29:13 2021, Anchal, Paco, Summary, ALS, Check for saturation in ITMX SOS Driver
|
Fri Jul 9 15:39:08 2021, Anchal, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat [Correction]
|
|
Message ID: 16164
Entry time: Thu May 27 11:03:15 2021
In reply to: 16161
Reply to this: 16168
|
Author: |
Anchal, Paco |
Type: |
Summary |
Category: |
ALS |
Subject: |
ALS Single Arm Noise Budget |
|
|
Here's an updated X ARM ALS noise budget.
Things to remember:
- Major mistake we were making earlier was that we were missing the step of clicking 'Set Phase UGF' before taking the measurement.
- Click the clear phase history just before taking measure.
- Make sure the IR beatnotes are less than 50 MHz (or the left half of HP8591E on water crate). The DFD is designed for this much beatnote frequency (from Gautum).
- We took this measurement with old IMC settings.
- We have saved a template file in users/Templates/ALS/ALS_outOfLoop_Ref_DQ.xml . This si same as ALS_outOfLoop_Ref.xml except we changed all channels to _DQ.
Conclusions:
- Attachment 1 shows the updated noisebudget. The estimated and measured RMS noise are very close to eachother.
- However, there is significant excess noise between 4 Hz and 200 Hz. We're still thinking on what could be the source of these.
- From 200 Hz to about 3 kHz, the beatnote noise is dominated by AUX residual frequency noise. This can be verified with page 2 of Attachment 2 where coherence between AUX PDH Error signal and BEATX signal is high.
- One mystery is how the measured beatnote noise is below the residual green laser noise above 3 kHz. Could this be just because the phase tracker can't measure noise above 3kHz?
- We have used estimated open loop transfer function for AUX from poles/zeros for uPDH box used (this was done months ago by me when I was working on ALS noise budget from home). We should verify it with a fresh OLTF measurement of AUX PDH loop. That's next on our list.
|
|
|