40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Thu Apr 29 11:51:27 2021, Anchal, Summary, LSC, Start of measuring IMC WFS noise contribution in ar cavity length noise
|
Thu Apr 29 15:11:33 2021, gautam, Update, CDS, RFM 
|
Thu Apr 29 17:43:16 2021, Koji, Update, CDS, RFM
|
Thu Apr 29 17:43:48 2021, Anchal, Update, CDS, F2A Filters double check
|
Fri Apr 30 00:20:30 2021, gautam, Update, CDS, F2A Filters double check
|
Thu Apr 29 17:51:19 2021, Anchal, Summary, LSC, Start of measuring IMC WFS noise contribution in arm cavity length noise
|
Fri Apr 30 00:18:40 2021, gautam, Summary, LSC, Start of measuring IMC WFS noise contribution in arm cavity length noise
|
Mon May 3 09:14:01 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise 
|
Mon May 3 17:28:58 2021, Anchal, Paco, Rana, Update, LSC, IMC WFS noise contribution in arm cavity length noise 
|
Tue May 4 11:43:09 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise  
|
Fri May 7 11:54:02 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise  
|
Mon May 10 18:19:12 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise, Corrections
|
Wed May 12 10:53:20 2021, Anchal, Paco, Update, LSC, PSL-IMC PDH Loop and XARM PDH Loop diagram
|
|
Message ID: 16104
Entry time: Fri Apr 30 00:18:40 2021
In reply to: 16101
|
Author: |
gautam |
Type: |
Summary |
Category: |
LSC |
Subject: |
Start of measuring IMC WFS noise contribution in arm cavity length noise |
|
|
This is the actuator calibration. For the error point calibration, you have to look at the filter in the calibration model. I think it's something like 8e-13m/ct for POX and similar for POY.
Quote: |
I calibrated the control arms signals by 2.44 nm/cts calibration factor directly picked up from 13984.
|
|