Entry  Fri Mar 19 19:15:13 2021, Jon, Update, CDS, c1auxey assembly 
    Reply  Thu Mar 25 12:11:53 2021, Yehonathan, Update, CDS, c1auxey assembly 20210325_115500_HDR.jpg20210325_123033.jpg
       Reply  Thu Mar 25 14:16:33 2021, gautam, Update, CDS, c1auxey assembly 
       Reply  Tue Mar 30 17:27:04 2021, Yehonathan, Update, CDS, c1auxey assembly 
          Reply  Wed Aug 25 11:48:48 2021, Yehonathan, Update, CDS, c1auxey assembly 
             Reply  Mon Sep 13 14:32:25 2021, Yehonathan, Update, CDS, c1auxey assembly ETMX_OSEMS_Noise.png
                Reply  Wed Sep 15 11:27:50 2021, Yehonathan, Update, CDS, c1auxey assembly c1auxex_Grounding_OSEM_comparison1.pdfc1auxex_Grounding_OSEM_comparison2.pdf
       Reply  Thu Apr 15 17:24:08 2021, Jon, Update, CDS, Updated c1auxey wiring plan C1AUXEY_Chassis_Feedthroughs_-_By_Connector.pdf
          Reply  Mon Apr 19 21:54:55 2021, Yehonathan, Update, CDS, Updated c1auxey wiring plan 
             Reply  Wed Apr 28 18:56:57 2021, Yehonathan, Jon, Update, CDS, Updated c1auxey wiring plan 
                Reply  Thu Apr 29 16:35:51 2021, Yehonathan, Update, CDS, Updated c1auxey wiring plan 
                   Reply  Thu Apr 29 19:55:45 2021, Yehonathan, Update, CDS, Updated c1auxey wiring plan 
                      Reply  Fri Apr 30 19:18:51 2021, Yehonathan, Jon, Update, CDS, Updated c1auxey wiring plan 
                         Reply  Mon May 3 20:36:46 2021, Yehonathan, Jon, Update, CDS, Updated c1auxey wiring plan 
                            Reply  Sun May 23 18:28:54 2021, Jon, Update, CDS, Opto-isolator for c1auxey 
                               Reply  Fri May 28 10:54:59 2021, Jon, Update, CDS, Opto-isolator for c1auxey optoisolator.jpeg
                                  Reply  Thu Jun 3 17:15:17 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_Isolator_NPN_Input.pngOptical_Isolator_NPN_Output.png
                                     Reply  Thu Jun 3 22:08:00 2021, Koji, Update, CDS, Opto-isolator for c1auxey 
                                        Reply  Fri Jun 4 14:49:23 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                           Reply  Fri Jun 4 17:46:25 2021, unYehonathan, Update, CDS, Opto-isolator for c1auxey 20210604_173420.jpg
                                              Reply  Sun Jun 6 16:33:47 2021, Jon, Update, CDS, BI channels on c1auxey 
                                                 Reply  Mon Jun 7 13:14:20 2021, Yehonathan, Update, CDS, BI channels on c1auxey 
                                                    Reply  Tue Jun 8 11:54:39 2021, Yehonathan, Update, CDS, BI channels on c1auxey 
                                           Reply  Sun Jun 6 03:02:14 2021, Koji, Update, CDS, Opto-isolator for c1auxey 
                                              Reply  Sun Jun 6 12:15:16 2021, Jon, Update, CDS, Opto-isolator for c1auxey 
                                              Reply  Sun Jun 6 15:59:51 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                                 Reply  Mon Jun 14 15:31:30 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                                    Reply  Tue Jun 15 21:48:55 2021, Koji, Update, CDS, Opto-isolator for c1auxey Acromag_RTS_BI_config.jpg
                                                       Reply  Wed Jun 16 17:24:29 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                                          Reply  Wed Jun 16 20:32:39 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey 
                                                             Reply  Fri Jul 9 18:35:32 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey 
                                                                Reply  Mon Jul 12 18:06:25 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey 
                                                                   Reply  Wed Jul 28 12:47:52 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey SUS-ETMY_SparePDMon0_NoRef.pngSUS-ETMY_SparePDMon0_Ref_WithGND.pngDifferentialOutputTest.png
                                                                      Reply  Wed Aug 11 12:06:40 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey SUS-ETMY_SparePDMon0_2.png
Message ID: 16037     Entry time: Thu Apr 15 17:24:08 2021     In reply to: 15962     Reply to this: 16052
Author: Jon 
Type: Update 
Category: CDS 
Subject: Updated c1auxey wiring plan 

I've updated the c1auxey wiring plan for compatibility with the new suspension electronics. Specifically it is based on wiring schematics for the new HAM-A coil driver (D1100117), satellite amplifier (D1002818), and HV bias driver (D1900163).

Changes:

  • The PDMon, VMon, CoilEnable, and BiasAdj channels all move from DB37 to various DB9 breakout boards.
  • The DB9 cables (x2) connecting the CoilEnable channels to the coil drivers must be spliced with the dewhitening switching signals from the RTS.
  • As suggested, I added five new BI channels to monitor the state of the CoilEnable switches. For lack of a better name, they follow the naming convention C1:SUS-ETMY_xx_ENABLEMon.

@Yehonathan can proceed with wiring the chassis.

Quote:

I finished prewiring the new c1auxey Acromag chassis (see attached pictures). I connected all grounds to the DIN rail to save some wiring. The power switches and LEDs work as expected.

I configured the DAQ modules using the old windows machine. I configured the gateway to be 192.168.114.1. The host machine still needs to be setup.

Next, the feedthroughs need to be wired and the channels need to be bench tested.

Attachment 1: C1AUXEY_Chassis_Feedthroughs_-_By_Connector.pdf  82 kB  Uploaded Thu Apr 15 18:42:41 2021  | Hide | Hide all
C1AUXEY_Chassis_Feedthroughs_-_By_Connector.pdf