- Added rough cts2mW calibration filters to the quadrants, see Attachment #1. The number I used is:
0.85 A/W * 500 V/A * 10 V/V * 1638.4 cts/V
(InGaAs responsivity) (RF transimpedance) (IQ demod conversion gain) (ADC calibration)
- Recovered FPMI locking. Once the arms are locked on POX / POY, I lock MICH using AS55_Q as a sensor and BS as an actuator with ~80 Hz UGF.
- Phased the digital demod phases such that while driving a sine wave in ETMX PIT, I saw it show up only in the "I" quadrant signals, see Attachment #2.
The idea is to use the FPMI config, which is more easily accessed than the PRFPMI, to set up some tests, measure some TFs etc, before trying to commission the more complicated optomechanical system. |