Mon Mar 8 12:01:02 2021, Paco, Anchal, Summary, training, Investigate how-to XARM locking
|
Mon Mar 8 12:40:35 2021, gautam, Summary, training, Investigate how-to XARM locking
|
Mon Mar 8 19:22:56 2021, rana, Summary, SUS, IMC suspension characterization
|
Thu Mar 11 08:13:24 2021, Paco, Anchal, Update, SUS, IMC First Free Swing Test failed due to typo, restarting now
|
Tue Mar 16 19:13:09 2021, Paco, Anchal, Update, SUS, First success in Input Matric Diagonalization 
|
Wed Mar 17 11:57:54 2021, Paco, Anchal, Update, SUS, Tested New Input Matrix for MC1 
|
Wed Mar 17 16:30:46 2021, Anchal, Update, SUS, Normalized Input Matrices plotted better than SURF students 
|
Thu Mar 18 09:18:49 2021, Paco, Anchal, Update, SUS, Testing of new input matrices with new data 
|
Thu Mar 18 12:46:53 2021, rana, Update, SUS, Testing of new input matrices with new data
|
Fri Mar 19 10:49:44 2021, Paco, Anchal, Update, SUS, Trying coil actuation balance
|
Mon Mar 22 11:57:21 2021, Paco, Anchal, Update, SUS, Trying coil actuation balance 
|
Mon Mar 22 15:10:00 2021, rana, Update, SUS, Trying coil actuation balance
|
Mon Mar 22 19:07:50 2021, Paco, Anchal, Update, SUS, Trying coil actuation balance 8x
|
Tue Mar 23 09:16:42 2021, Paco, Anchal, Update, Computers, Power cycled C1PSL; restored C1PSL 
|
Wed Mar 24 09:23:49 2021, Paco, Update, SUS, MC3 new Input Matrix
|
Thu Mar 25 11:46:31 2021, Paco, Anchal, Update, SUS, MC2 Coil Balancing updates   
|
Thu Mar 25 15:31:24 2021, gautam, Update, IOO, WFS servos
|
Fri Mar 26 11:54:37 2021, Paco, Anchal, Update, SUS, MC2 Coil Balancing updates
|
Mon Mar 29 10:44:51 2021, Paco, Anchal, Update, SUS, MC2 Coil Balancing updates 
|
Mon Mar 29 17:11:54 2021, gautam, Update, SUS, MC2 Coil Balancing updates
|
Mon Mar 29 17:34:52 2021, rana, Update, SUS, MC2 Coil Balancing updates
|
Wed Mar 31 22:58:32 2021, Anchal, Paco, Update, SUS, MC2 Coil Balancing Test
|
Thu Apr 1 13:56:49 2021, Anchal, Paco, Update, SUS, MC2 Coil Balancing Test Results  
|
Thu Apr 1 18:01:06 2021, Anchal, Paco, Update, SUS, MC2 Coil Balancing Test Results Success??  
|
Thu Apr 1 18:48:45 2021, gautam, Update, SUS, MC2 Coil Balancing Test Results Success??
|
Thu Apr 1 21:13:54 2021, Anchal, Update, SUS, Matrix results, new measurement set to trigger
|
Fri Apr 2 14:51:20 2021, Anchal, Update, SUS, Bug found, need to redo the balancing
|
Tue Apr 6 18:46:36 2021, Anchal, Paco, Update, SUS, Updates on recent efforts
|
Wed Apr 7 17:38:51 2021, Anchal, Update, SUS, Trying to uncouple only PIT and YAW first    
|
Thu Apr 8 17:04:43 2021, Anchal, Paco, Update, SUS, First Successful Coil Balancing  
|
Fri Apr 9 13:13:00 2021, Anchal, Paco, Update, SUS, Faster coil balancing 
|
Fri Apr 9 17:41:12 2021, rana, Update, SUS, Faster coil balancing
|
Mon Apr 12 10:07:35 2021, Anchal, Update, SUS, What's F2A??
|
Tue Apr 13 09:51:22 2021, rana, Update, SUS, What's F2A??
|
Wed Apr 14 17:53:38 2021, Anchal, Update, SUS, Plan for calculating filter banks for output matrix aka F2A aka F2P
|
Thu Apr 15 11:41:43 2021, Anchal, Update, SUS, Proposed filters for output matrix aka F2A aka F2P
|
Fri Apr 16 11:36:36 2021, Anchal, Paco, Update, SUS, Tested proposed filters for POS colum in MC2 output matrix    
|
Fri Apr 16 15:47:58 2021, rana, Update, SUS, Tested proposed filters for POS colum in MC2 output matrix
|
Mon Apr 19 12:18:19 2021, Anchal, Paco, Update, SUS, Tested proposed filters for POS colum in MC2 output matrix  
|
Tue Apr 20 10:52:49 2021, Anchal, Paco, Update, SUS, AC gain coil output balancing for IMC   
|
Tue Apr 20 18:19:30 2021, Anchal, Update, SUS, MC2 coil balanced at DC  
|
Wed Apr 21 11:38:27 2021, Anchal, Paco, Update, SUS, MC2 Damping Gains Optimized 
|
Wed Apr 21 15:50:01 2021, Anchal, Paco, Update, SUS, MC2 Suspension Optimization summary
|
Wed Apr 21 18:49:29 2021, rana, Update, SUS, MC2 Suspension Optimization summary
|
Thu Apr 22 08:50:21 2021, Anchal, Update, SUS, MC2 Suspension Optimization summary
|
Mon Apr 26 18:55:39 2021, Anchal, Paco, Update, SUS, MC2 F2A Filters Tested  
|
Tue Apr 27 10:05:28 2021, Anchal, Paco, Update, SUS, MC1 and MC3 F2A Filters Tested   
|
Wed Apr 28 10:56:10 2021, Anchal, Paco, Update, SUS, IMC Filters diagnosed 7x
|
Wed Apr 28 17:09:11 2021, Anchal, Update, SUS, Tuned Suspension Parameters uploaded for long term behavior monitoring
|
Thu Apr 29 10:52:56 2021, Anchal, Update, SUS, IMC Trans QPD and WFS loops step response test 
|
Thu Apr 29 18:53:33 2021, Anchal, Update, SUS, IMC Suspension Damping Gains Test  
|
Mon May 3 16:24:14 2021, Anchal, Update, SUS, IMC Suspension Damping Gains Test Repeated with IMC unlocked  
|
Wed May 5 09:04:47 2021, Anchal, Update, SUS, New IMC Suspension Damping Gains uploaded for long term testing
|
Wed May 12 11:45:13 2021, Anchal, Paco, Summary, SUS, New IMC Settings are miserable  
|
Wed Jun 2 09:43:30 2021, Anchal, Paco, Summary, SUS, IMC Settings characterization   
|
Wed Jun 2 16:20:59 2021, Anchal, Paco, Summary, SUS, IMC Suspension gains reverted to old values
|
Sat May 15 14:54:24 2021, gautam, Update, SUS, IMC settings reverted
|
Thu May 20 10:35:57 2021, Anchal, Update, SUS, IMC settings reverted
|
Thu Apr 22 12:17:23 2021, Anchal, Paco, Update, SUS, MC1 and MC3 Suspension Optimization Summary
|
Thu Apr 22 14:22:39 2021, gautam, Update, SUS, Settings restored
|
Thu Apr 22 15:36:54 2021, Anchal, Update, SUS, Settings restored
|
Thu Apr 22 17:04:17 2021, gautam, Update, SUS, Settings restored
|
Sat Apr 10 10:07:47 2021, rana, Update, SUS, Faster coil balancing
|
Fri Apr 2 15:22:54 2021, gautam, Update, SUS, Matrix results, new measurement set to trigger
|
Sun Mar 28 14:16:25 2021, Anchal, Summary, SUS, MC3 new Input Matrix not providing stable loop
|
Mon Mar 29 17:07:17 2021, gautam, Summary, SUS, MC3 new Input Matrix not providing stable loop
|
Thu Mar 18 13:12:39 2021, gautam, Update, Computer Scripts / Programs, Omnigraffle vs draw.io
|
|
Message ID: 15877
Entry time: Mon Mar 8 12:01:02 2021
Reply to this: 15878
15881
|
Author: |
Paco, Anchal |
Type: |
Summary |
Category: |
training |
Subject: |
Investigate how-to XARM locking |
|
|
[Paco, Anchal]
- Started zoom stream; thanks to whoever installed it!
- Spent some time trying to understand how anything we did last thursday lead to the sensing matrix change, but still cannot figure it out.
- Tracking back on our actions, at ~10:30 we ran burt Restore with the 08:19/.*snap and in lack of a better suspect, we blame it on that action for now.
# ARM locking??
- Reading (not running) the scripts/XARM/lockXarm.py script and try to understand the workflow. It is pretty confusing that the result was to lock Yarm last time.
- It looks like this script was a copy of lockYarm.py, and was never updated (there's a chance we ran it for the first time last thursday)
- *Is there a script to lock the Arms?* Or should we write one? To write one, we first attempt a manual procedure;
1. No need to change RFPD InMTRX
2. All filters inputs / outputs are enabled
3. Outputs from XARM and YARM in the Output matrix are already going to ETMX and ETMY
- Maybe we can have the ARM lock engage by changing the MC directly?
4. Change C1:SUS-MC2_POS_OFFSET from -38 to -0, and enable C1:SUS-MC2_POS_OFFSET_ON
5. Manually scan MC2_POS_OFFSET to 250 (nothing happens), then -250, then back to -38 (WFS1 PIT and YAW changed a little, but then returned to their nominal values)
- Or maybe we need to provide the right gain...
6. Disabled C1:SUS-MC2_POS_OFFSET_ON (back to nominal state)
7. Look into manually changing C1:LSC-XARM_GAIN;
From the command line using python:
>> import epics
>> ch_name = 'C1:LSC-XARM_GAIN'
>> epics.caput(ch_name, 0.155) # nominal = 0.150
- Could be unrelated, but we noted a slow spike on C1:PSL-FSS_PCDRIVE (definitely from before we changed anything)
- Still nothing is happening
8. Changed the gain to 0.175, then back to 0.150, no effect... then 0.2, 0.3 ...
- Stop and check SUS_Watchdogs (should not have changed?) and everything remains nominal
- Revert all changes symmetrically.
- Could we have missed enabling FM1?
- Briefly lost MC lock, but it came back on its own (probably unrelated)
- Wrap it up for the day. In summary; no harm done to our knowledge. |