Sat Feb 20 16:20:33 2021, gautam, Update, General, Housekeeping + PRMI char   
|
Sun Feb 21 20:51:21 2021, rana, Update, General, Housekeeping + PRMI char
|
Tue Feb 23 00:10:05 2021, gautam, Update, General, Demod char part 1 
|
Wed Feb 24 11:53:24 2021, gautam, Update, General, Demod char part 2 
|
Wed Feb 24 12:11:08 2021, gautam, Update, General, Demod char part 3   
|
Thu Feb 25 16:50:53 2021, gautam, Update, General, PRMI sensing matrix
|
Thu Feb 25 20:37:49 2021, gautam, Update, General, Setting modulation frequency and checking IMC offset 
|
|
Message ID: 15844
Entry time: Thu Feb 25 16:50:53 2021
In reply to: 15840
Reply to this: 15845
|
Author: |
gautam |
Type: |
Update |
Category: |
General |
Subject: |
PRMI sensing matrix |
|
|
After all the work at the LSC rack over the last couple of days, I re-locked the PRMI (ETMs misaligned), and measured the sensing matrix once again. The PRMI was locked using 1f error signals, with AS55_Q as the MICH sensor and REFL11_I as the PRCL sensor. As shown in Attachment #1, the situation has not changed, there is still no separation between the DoFs in the REFL signals. I will measure the MC lock point offset using the error point dither technique today to see if there is something there. |
|
|