40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Jan 20 18:07:09 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements before modifications 7x
    Reply  Mon Jan 25 18:18:04 2021, Anchal, Summary, BHD, Satellite Amplifier Transfer Functions and noise D1002818_S2100029_TFandNoiseSpectrum.pdfD1002818_Testing.zip
       Reply  Wed Feb 3 16:27:19 2021, Anchal, Summary, BHD, Satellite Amplifier Transfer Functions and noise After modifications D1002818_S2100029_TFAfterChanges.pdfD1002818_S2100029_OutputNoiseSpecAfterChanges.pdfD1002818_S2100029_InputRefferedNoiseSpecAfterChanges.pdfD1002812_S2100029_After_Modifications_Feb3.jpgAfterChanges.zip
          Reply  Wed Feb 10 11:45:59 2021, Anchal, Summary, BHD, Satellite Amplifier Very Low frequency noise After modifications 6x
             Reply  Wed Feb 10 15:07:50 2021, Anchal, Summary, BHD, Satellite Amplifier Output Offset measurements 
                Reply  Wed Feb 10 15:25:45 2021, gautam, Summary, BHD, Satellite Amplifier Output Offset measurements 
                   Reply  Wed Feb 10 17:18:03 2021, Koji, Summary, BHD, Satellite Amplifier Output Offset measurements 
             Reply  Thu Feb 11 11:10:05 2021, Anchal, Summary, BHD, Satellite Amplifier Very Low frequency noise After modifications AfterChangesSpectrum_AC.zipD1002818_S2100029_OutputNoiseSpecAfterChanges.pdfD1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdfD1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf
                Reply  Thu Feb 11 16:58:52 2021, rana, Summary, BHD, Satellite Amplifier Very Low frequency noise After modifications 
    Reply  Thu Jan 28 12:53:14 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements before modifications D1100117_S2100027_Current_Noise_Spectrum.pdfD1100117_S2100027_Voltage_Noise_Spectrum.pdfD1100117_S2100028_Current_Noise_Spectrum.pdfD1100117_S2100028_Voltage_Noise_Spectrum.pdfSpectrumMeasurement.zip
       Reply  Thu Jan 28 18:04:55 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements After modifications D1100117_S2100027_TF.pdfAfterChanges.zip
          Reply  Thu Jan 28 21:44:45 2021, gautam, Summary, BHD, HAM-A Coil Driver measurements After modifications 
             Reply  Fri Jan 29 15:39:30 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements After modifications TF and Noise S2100027 D1100687_S2100027_After_Modifications_Jan28.jpgD1100117_S2100027_TF.pdfD1100117_S2100027_Voltage_Noise_Spectrum.pdfD1100117_S2100027_Current_Noise_Spectrum.pdfAfterChanges.zip
                Reply  Tue Feb 2 11:57:46 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements After modifications TF and Noise S2100028 D1100687_S2100028_After_Modifications_Feb01_2021.jpgD1100117_S2100028_TF.pdfD1100117_S2100028_Voltage_Noise_Spectrum.pdfD1100117_S2100028_Current_Noise_Spectrum.pdfAfterChanges.zip
Message ID: 15803     Entry time: Thu Feb 11 11:10:05 2021     In reply to: 15797     Reply to this: 15804
Author: Anchal 
Type: Summary 
Category: BHD 
Subject: Satellite Amplifier Very Low frequency noise After modifications 

Here is a proper measurement for PD transimpedance amplifier circuit in the Satellite amplifier box D1002818 S2100029. The input from rear DB25 connector was left open and measurement was taken with AC coupling with correction by the AC coupling transfer function (Zero at 0, pole at 160 mHz). I have calculated the input referred displacement noise by calculating the conversion factor of OSEM in A/m. From 40m/12470, old conversion factor of OSEM to output of sat amplifier was 1.6 V/mm. then, the transimpedance was 39.2 kOhm, so that must mean a conversion factor of 1.6e3/39.2 A/m. This I scaled with increased drive current by factor of 35/25 as mentioned in this document. The final conversion factor turned out to be around 57 mA / m. If someone finds error in this, please let me know.

There is excess noise in the low-frequency region below 5-6 Hz. If people think I should make a measurement of amplified noise to go further away from the instrument noise floor, let me know.

Attachment 1: AfterChangesSpectrum_AC.zip  10.764 MB
Attachment 2: D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf  183 kB  Uploaded Thu Feb 11 17:02:26 2021  | Hide | Hide all
D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseSpecAfterChanges.pdf
Attachment 3: D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf  277 kB  Uploaded Thu Feb 11 17:02:53 2021  | Hide | Hide all
D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdf
Attachment 4: D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf  276 kB  Uploaded Thu Feb 11 17:03:10 2021  | Hide | Hide all
D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf D1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf