Last week and this week I've been working on the characterization of the Q3000 QPDs. The QPDs were named 81, 82, 83, and 94.
- Dark current [OMC LAB ELOG 402]: All the segments looked similar and acceptable except for the seg1 of #82. It has a smaller reverse breakdown voltage (~6V) but even this is an acceptable level.
- Impedance [OMC LAB ELOG 403]: All the segments showed a ~300pF junction capacitance with no reverse bias. This looks quite normal.
- Dark noise [OMC LAB ELOG 404]: All the segments showed ~5pA/rtHz dark noise above 1Hz.
My recommendation is to use #81 and #84 as they have similar dark current characteristics between the segments. But basically, all the QPDs look fine.
The actual junction capacitance and the RF dark noise should be characterized by the actual WFS head circuit.
The QPD packages were labeled and returned to Gautam to be implemented in the WFS heads.
gautam: S/N #84 was installed as the AS WFS QPD. The remaining 3 are stored in the clean cabinet at EX (where the rest of the RF photodiodes are). |