Thu Jan 10 19:59:08 2019, Jon, Update, VAC, Vac System Running Normally on Turbo Pumps  
|
Fri Jan 25 16:14:51 2019, gautam, Update, VAC, Vacuum interlock code, N2 warning
|
Wed Mar 25 15:15:55 2020, gautam, Update, VAC, Vacuum interlock code, N2 warning update
|
|
Message ID: 15283
Entry time: Wed Mar 25 15:15:55 2020
In reply to: 14419
|
Author: |
gautam |
Type: |
Update |
Category: |
VAC |
Subject: |
Vacuum interlock code, N2 warning update |
|
|
The email address in the N2 checking script wasn't right - I now updated it to email the 40m list if the sum of reserve tank pressures fall below 800 PSI. The checker itself is only run every 3 hours (via cron on c1vac).
Quote: |
I reset the remote of this git repo to the 40m version instead of Jon's personal one, to ensure consistency between what's on the vacuum machine and in the git repo. There is now a N2 checker python mailer that will email the 40m list if all the tank pressures are below 600 PSI (>12 hours left for someone to react before the main N2 line pressure drops and the interlocks kick in). For now, the script just runs as a cron job every 3 hours, but perhaps we should integrate it with the interlock process
|
|