Message ID: 15202     Entry time: Mon Feb 10 10:07:20 2020
Author: gautam 
Type: Update 
Category: PSL 
Subject: PMC re-locked 

I found the PMC unlocked this morning. It was re-locked using the usual procedure. I feel like this has been happening more frequently in the last month than before. In the past, the cause seems to have been the PZT voltage drifting too close to one of the rails - however, in this case, it looks like an IMC unlock event is what triggered the PMC lockloss (admittedly the PZT voltage was somewhat close to the rail). It would be good if someone can re-connect the PMC Transmission photodiode, it was a useful diagnostic channel we had working fine before the ringdowns started.

I also tweaked the input pointing into the PMC and ran the WFS DC offset relief script.