40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Dec 27 16:26:11 2019, Yehonathan, Update, PSL, PMC Linewidth measurement PMCTransmissionSpectra.pdf
    Reply  Wed Jan 1 14:14:00 2020, Yehonathan, Update, PSL, PMC Linewidth measurement 150MHzSideBandCreation.pdfLinewidthMeasurment.pdf
       Reply  Wed Jan 15 15:04:31 2020, gautam, Update, PSL, PMC Linewidth measurement 
          Reply  Fri Jan 17 22:11:19 2020, Yehonathan, Update, PSL, Ringdown measurements 20200117_192455.jpg20200117_192448.jpg
             Reply  Sat Jan 25 11:54:42 2020, Yehonathan, Update, PSL, Ringdown measurements Input_beam_path.jpgReflected_Beam_Path.jpgTransmitted_Beam_Path.jpgPMCRingdownNormalizedRawdata.pdfTransPDFits.pdf
                Reply  Mon Jan 27 21:35:06 2020, Yehonathan, Update, PSL, Ringdown measurements IMC_REFL_Beam_Path.jpg
                   Reply  Mon Jan 27 21:48:49 2020, gautam, Update, PSL, Ringdown measurements elevatedSeis.pdf
                      Reply  Tue Jan 28 16:01:17 2020, gautam, Update, IOO, IMC WFS servos stable again QPD_recenter.png
                         Reply  Tue Jan 28 20:51:37 2020, Yehonathan, Update, IOO, IMC WFS servos stable again 
                            Reply  Wed Jan 29 12:40:24 2020, Yehonathan, Update, IOO, IMC Ringdowns preliminary data analysis IMCRingdownNormalizedRawdata.pdfIMCTransPDFits.pdf
                               Reply  Mon Feb 3 13:54:10 2020, Yehonathan, Update, IOO, IMC Ringdowns extended data analysis IMCTransReflAnalysis_anotated.pdf
                                  Reply  Wed Feb 5 21:13:17 2020, Yehonathan, Update, IOO, IMC Ringdowns extended data analysis 
                                     Reply  Sat Feb 15 12:56:24 2020, Yehonathan, Update, IOO, IMC Transfer function measurement IMCFrequencyResponse.pdfIMCRingdownNormalizedRawdata.pdfIMCREFLPDFits.pdf
Message ID: 15170     Entry time: Tue Jan 28 20:51:37 2020     In reply to: 15165     Reply to this: 15175
Author: Yehonathan 
Type: Update 
Category: IOO 
Subject: IMC WFS servos stable again 

I resume my IMC ringdown activities now that the IMC is aligned again.

To avoid any accidental misalignments Gautam turned off all the inputs to the WFS servo.

I set up a PD and a lens as in attachment 1 (following Gautam's setup).

I connect the REFL, TRANS and INPut PDs to the oscilloscope.

I connect a Siglent function generator to the AOM driver. I try to shut off the light to the IMC using 1V DC waveform and pressing the output button manually. However, it produced heavily distorted step function in the PMC trans PD.

I use a square wave with a frequency of 20mHz instead with an amplitude of 0.5V offset of 0.25V and dutycycle of 1% so there will be minimal wasted time in the off state. I get nice ringdowns (attachment 2) - forgot to take pictures. The autolocker slightly misaligns the M2 every time it is acting, so I manually align it everytime the IMC gets unlocked.

Data analysis will come later.

I remove the PD and lens and reenable the WFS servo inputs. The IMC locks easily. The WFS outputs are very different than 0 now though.