40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Dec 27 16:26:11 2019, Yehonathan, Update, PSL, PMC Linewidth measurement PMCTransmissionSpectra.pdf
    Reply  Wed Jan 1 14:14:00 2020, Yehonathan, Update, PSL, PMC Linewidth measurement 150MHzSideBandCreation.pdfLinewidthMeasurment.pdf
       Reply  Wed Jan 15 15:04:31 2020, gautam, Update, PSL, PMC Linewidth measurement 
          Reply  Fri Jan 17 22:11:19 2020, Yehonathan, Update, PSL, Ringdown measurements 20200117_192455.jpg20200117_192448.jpg
             Reply  Sat Jan 25 11:54:42 2020, Yehonathan, Update, PSL, Ringdown measurements Input_beam_path.jpgReflected_Beam_Path.jpgTransmitted_Beam_Path.jpgPMCRingdownNormalizedRawdata.pdfTransPDFits.pdf
                Reply  Mon Jan 27 21:35:06 2020, Yehonathan, Update, PSL, Ringdown measurements IMC_REFL_Beam_Path.jpg
                   Reply  Mon Jan 27 21:48:49 2020, gautam, Update, PSL, Ringdown measurements elevatedSeis.pdf
                      Reply  Tue Jan 28 16:01:17 2020, gautam, Update, IOO, IMC WFS servos stable again QPD_recenter.png
                         Reply  Tue Jan 28 20:51:37 2020, Yehonathan, Update, IOO, IMC WFS servos stable again 
                            Reply  Wed Jan 29 12:40:24 2020, Yehonathan, Update, IOO, IMC Ringdowns preliminary data analysis IMCRingdownNormalizedRawdata.pdfIMCTransPDFits.pdf
                               Reply  Mon Feb 3 13:54:10 2020, Yehonathan, Update, IOO, IMC Ringdowns extended data analysis IMCTransReflAnalysis_anotated.pdf
                                  Reply  Wed Feb 5 21:13:17 2020, Yehonathan, Update, IOO, IMC Ringdowns extended data analysis 
                                     Reply  Sat Feb 15 12:56:24 2020, Yehonathan, Update, IOO, IMC Transfer function measurement IMCFrequencyResponse.pdfIMCRingdownNormalizedRawdata.pdfIMCREFLPDFits.pdf
Message ID: 15106     Entry time: Fri Dec 27 16:26:11 2019     Reply to this: 15109
Author: Yehonathan 
Type: Update 
Category: PSL 
Subject: PMC Linewidth measurement 

I try to measure the linewidth of the PMC by ramping the PMC PZT. 

I do it by connecting a triangular shape signal to FP Test 1 on the PMC servo front panel (I know, it is probably better to connect it to DC EXT. next time.) and turn the servo gain to a minimum.

Attachment 1 shows the PMC transmission PD as the PZT is swept with the EOM connected and when it is disconnected. It shows the PMC over more than 1 free spectral range.

For some reason, I cannot seem to be able to find the 35MHz sidebands which I want to use to calibrate the PZT scan. I made sure that the EOM is driven by a 35MHz signal using the scope. I also made sure that the PMC cannot to lock without the EOM connected.

I am probably doing something silly.

Attachment 1: PMCTransmissionSpectra.pdf  97 kB  Uploaded Sat Dec 28 11:05:25 2019  | Hide | Hide all
PMCTransmissionSpectra.pdf
ELOG V3.1.3-