Message ID: 15050     Entry time: Tue Nov 26 18:16:08 2019     In reply to: 15049     Reply to this: 15051
Author: rana 
Type: Update 
Category: LSC 
Subject: POX / POY calibration 

...maybe the opto-mechanical CARM plant is changing as a function of the CARM offset...

Quote:

Even assuming 50% error in the calibration factors, it's hard to explain the swing of TRX/TRY when the CARM offset is brought to zero.

  • The increase in (admittedly in-loop) CARM noise as the offset is reduced still seems to me to be correlated with the buildup of IR power in the arm cavities.