40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Oct 8 18:35:09 2019, gautam, Update, LSC, Locking prep 
    Reply  Wed Oct 9 18:15:26 2019, gautam, Update, LSC, PRMI 3f locking PRMI_1f.png
       Reply  Wed Oct 9 22:02:58 2019, gautam, Update, LSC, REFL55 whitening issue 
          Reply  Thu Oct 10 01:12:56 2019, gautam, Update, LSC, Locking studies PRFPMIstudies.png
             Reply  Wed Oct 16 11:42:17 2019, gautam, Update, LSC, Poor separation of PRCL/MICH in 3f signals sensMat.pdf
                Reply  Thu Oct 17 11:19:28 2019, gautam, Update, LSC, Locking activity last night MICH2PRCLnulling.pdfconsolidatedSensingMatrices.pdf.zip
                   Reply  Tue Oct 22 00:52:27 2019, gautam, Update, LSC, Locking updates PRMI3f_ALS_Oct21sensMat.pdf
                      Reply  Tue Oct 29 01:24:45 2019, gautam, Update, LSC, More locking updates POP22_feature.pngPRMI3f_ALS_Oct21sensMat.pdf
                         Reply  Tue Oct 29 15:13:19 2019, gautam, Update, LSC, More locking updates PRMI_lockloss.png
                            Reply  Wed Oct 30 01:27:00 2019, gautam, Update, LSC, More locking updates POP_ASS.png
                               Reply  Wed Oct 30 11:53:41 2019, gautam, Update, LSC, MICH loop shape tuning MICH_ELP.pdfMICH_ELP_TFs.pdf
                               Reply  Wed Oct 30 17:08:40 2019, gautam, Update, LSC, POP22 investigation POP22anomaly.pdf
                                  Reply  Mon Nov 4 15:29:47 2019, gautam, Update, LSC, POP signal path POPchain.pdf
                                     Reply  Mon Nov 4 16:06:58 2019, gautam, Update, LSC, POP optical path IMG_8027.JPGlensRealignment.jpgPOPrealigned.pngPOPdither.pngPRMfixed.png
                                        Reply  Tue Nov 5 11:52:27 2019, gautam, Update, LSC, Locking notes PRMI3f_ALS_Nov4sensMat.pdf
                                           Reply  Wed Nov 6 02:08:48 2019, gautam, Update, LSC, Locking updates 
                                              Reply  Wed Nov 6 17:05:45 2019, gautam, Update, LSC, CARM calibration CARMscan.pdfCARMcalib.pdfscan.hdf5.zip
                                              Reply  Wed Nov 6 17:45:34 2019, gautam, Update, LSC, ~ 
                                                 Reply  Mon Nov 18 21:04:38 2019, gautam, Update, LSC, Locking - some ideas 
                                                    Reply  Wed Nov 20 01:07:18 2019, gautam, Update, LSC, Locking - progress PRFPMI.png
                                                       Reply  Wed Nov 20 12:14:17 2019, gautam, Update, LSC, Locking - progress TRX_TRY_comparison.pdf
                                                          Reply  Wed Nov 20 17:20:24 2019, Yehonathan, Update, LSC, QPD Investigation 
                                                             Reply  Wed Nov 20 17:52:00 2019, gautam, Update, LSC, QPD MEDM screen update newLookQPD.pngTRX_TRY_comparison.pdfIMG_8186.JPG
                                                                Reply  Wed Nov 20 21:29:28 2019, gautam, Update, LSC, PRG ~13 PRG13.pdf
                                                                   Reply  Fri Nov 22 00:54:14 2019, gautam, Update, LSC, locking notes ALSnoiseIncrease.pdf
                                                                      Reply  Mon Nov 25 19:11:22 2019, gautam, Update, LSC, ALS noise re-look ALSnoise.pdf
                                                                         Reply  Wed Nov 27 23:24:01 2019, gautam, Update, LSC, No shaking but no inspiration either 
                                                                            Reply  Sun Dec 1 13:38:48 2019, rana, Update, LSC, No shaking but no inspiration either 
                                                                               Reply  Tue Dec 3 00:03:57 2019, gautam, Update, LSC, Green ALS also shows elevated noise with high arm buildup ALSnoiseIncrease_greenBeat.pdf
                                                                                  Reply  Wed Dec 4 19:54:27 2019, gautam, Update, LSC, A look to the past DRFPMI_2016March.pdf
                                                                                     Reply  Thu Dec 5 01:54:39 2019, gautam, Update, LSC, Partial CM board path engaged AOengaged.pdf
                                                                                        Reply  Fri Dec 6 00:02:48 2019, gautam, Update, LSC, What is the correct way to set the 3f offsets? 
                                     Reply  Fri Nov 15 12:08:04 2019, gautam, Update, LSC, POPDC whitening board 
                         Reply  Tue Oct 29 17:40:48 2019, gautam, Update, LSC, More locking updates broadSweep.pdfzoomSweep.pdf
Message ID: 15046     Entry time: Mon Nov 25 19:11:22 2019     In reply to: 15045     Reply to this: 15056
Author: gautam 
Type: Update 
Category: LSC 
Subject: ALS noise re-look 

I re-checked the ALS noise in the following configurations:

  • PRM is misaligned.
  • Michelson is not locked.
  • TRX/TRY is maintained at ~1.
  1. Arm lengths are controlled using POX/POY as a sensor, and the ETMs as actuators [orange traces in Attachment #1].
    • EX laser frequency is locked to the arm cavity length using the end PDH servo.
    • ALS beat note frequency fluctuations are read out using the calibrated DFD channels.
    • In this config, the DFD outputs are the out-of-loop sensor.
  2. Arm lengths are controlled using the ALS beat frequencies as a sensors [blue traces in Attachment #1]
    • The control is no longer in the XARM/YARM basis, but in the CARM/DARM basis.
    • The CARM actuator is MC2, the DARM actuator is an admixture of the ETMs (equal magnitude of output matrix element, opposite sign).
    • The calibrated POX/POY photodiodes are used as the out-of-loop sensor in this config.

The RMS noise sensed by POX/POY is ~20pm, which is somewhat higher than the best I've seen (maybe the arms are moving more at the time of measurement or the AUX PDH loops need a bit of touching up). But the orange traces in the top row of Attachment #1 are already ~x2 better than the equivalent traces from when we were using the green beams to make the beats. So it's hard to explain the 0-300 fluctuations in the arm powers when the CARM offset is reduced to 0 - i.e. the ALS noise is becoming worse as I reduce the CARM offset (= have more circulating power compared to the conditions of this test). I assume the transmission is Lorentzian, in which case even if we have 5x the CARM linewidth worth of ALS noise, we should see the arm power fluctuate between ~10 and 300. 

* I notice that a big jump in the RMS sensed by POX/POY comes from the 24 Hz peak, which is presumably the Roll mode coupling to length - maybe a ResG can make the situation better. The high frequency noise can also be probably rolled off better.

Attachment 1: ALSnoise.pdf  96 kB  | Hide | Hide all