Tue Oct 8 18:35:09 2019, gautam, Update, LSC, Locking prep
|
Wed Oct 9 18:15:26 2019, gautam, Update, LSC, PRMI 3f locking
|
Wed Oct 9 22:02:58 2019, gautam, Update, LSC, REFL55 whitening issue
|
Thu Oct 10 01:12:56 2019, gautam, Update, LSC, Locking studies
|
Wed Oct 16 11:42:17 2019, gautam, Update, LSC, Poor separation of PRCL/MICH in 3f signals
|
Thu Oct 17 11:19:28 2019, gautam, Update, LSC, Locking activity last night 
|
Tue Oct 22 00:52:27 2019, gautam, Update, LSC, Locking updates
|
Tue Oct 29 01:24:45 2019, gautam, Update, LSC, More locking updates 
|
Tue Oct 29 15:13:19 2019, gautam, Update, LSC, More locking updates
|
Wed Oct 30 01:27:00 2019, gautam, Update, LSC, More locking updates
|
Wed Oct 30 11:53:41 2019, gautam, Update, LSC, MICH loop shape tuning 
|
Wed Oct 30 17:08:40 2019, gautam, Update, LSC, POP22 investigation
|
Mon Nov 4 15:29:47 2019, gautam, Update, LSC, POP signal path
|
Mon Nov 4 16:06:58 2019, gautam, Update, LSC, POP optical path    
|
Tue Nov 5 11:52:27 2019, gautam, Update, LSC, Locking notes
|
Wed Nov 6 02:08:48 2019, gautam, Update, LSC, Locking updates
|
Wed Nov 6 17:05:45 2019, gautam, Update, LSC, CARM calibration  
|
Wed Nov 6 17:45:34 2019, gautam, Update, LSC, ~
|
Mon Nov 18 21:04:38 2019, gautam, Update, LSC, Locking - some ideas
|
Wed Nov 20 01:07:18 2019, gautam, Update, LSC, Locking - progress
|
Wed Nov 20 12:14:17 2019, gautam, Update, LSC, Locking - progress
|
Wed Nov 20 17:20:24 2019, Yehonathan, Update, LSC, QPD Investigation
|
Wed Nov 20 17:52:00 2019, gautam, Update, LSC, QPD MEDM screen update  
|
Wed Nov 20 21:29:28 2019, gautam, Update, LSC, PRG ~13
|
Fri Nov 22 00:54:14 2019, gautam, Update, LSC, locking notes
|
Mon Nov 25 19:11:22 2019, gautam, Update, LSC, ALS noise re-look
|
Wed Nov 27 23:24:01 2019, gautam, Update, LSC, No shaking but no inspiration either
|
Sun Dec 1 13:38:48 2019, rana, Update, LSC, No shaking but no inspiration either
|
Tue Dec 3 00:03:57 2019, gautam, Update, LSC, Green ALS also shows elevated noise with high arm buildup
|
Wed Dec 4 19:54:27 2019, gautam, Update, LSC, A look to the past
|
Thu Dec 5 01:54:39 2019, gautam, Update, LSC, Partial CM board path engaged
|
Fri Dec 6 00:02:48 2019, gautam, Update, LSC, What is the correct way to set the 3f offsets?
|
Fri Nov 15 12:08:04 2019, gautam, Update, LSC, POPDC whitening board
|
Tue Oct 29 17:40:48 2019, gautam, Update, LSC, More locking updates 
|
|
Message ID: 15010
Entry time: Mon Nov 4 16:06:58 2019
In reply to: 15009
Reply to this: 15012
|
Author: |
gautam |
Type: |
Update |
Category: |
LSC |
Subject: |
POP optical path |
|
|
I did some re-alignment of the POP beam on the IX in air table. Here are the details:
- Attachment #1 - optical layout.
- With the PRC locked with the carrier resonant (no arm cavities), there is ~300uW of DC power incident on the Thorlabs PDA10CF, which serves as POP22, POP110 and POPDC photosensor.
- See this elog for the signal paths.
- On a scope, this corresponded to ~1.8 V DC of voltage. This is in good agreement with the expected transimpedance gain of 10 kOhms and responsivity of ~0.65 A/W given on the datasheet.
- This is also in agreement with the ~6000 ADC counts I see in the CDS system (although there are large fluctuations).
- These was significant misalignment of the beam on this photodiode at some point:
- Previously, I had used the CDS system to walk the beam on thde photodiode to try and maximize the power.
- Today I took a different approach - triggered the MICH and PRCL loops on REFLDC (instead of the usual POPDC / POP22) so I could freely block the beam.
- I found that there is a fast (f=35mm) lens to make the beam small enough for the PDA10CF. The beam was somewhat mis-centered on this strongly curved optic, and I suspect it was amplifying small misalignments. Anyway it is much better centered now (see Attachment #2) and I have a much stronger POPDC signal (by a factor of ~2-3, see Attachment #3).
- The ASS dither alignment now shows much more consistent behavior - minimizing REFLDC maximises POPDC, see Attachment #4.
- I took this opportunity to take some spectra/time-series of the PD output with the interferometer in this configuration.
Tangentially related to this work - I took the nuclear option and did a hard reboot of the c1susaux Acromag crate on Sunday to fix the EPICS issue - it seems to be gone for now, see Attachment #5. |
|
|
|
|
|
|
|
|
|