Tue Oct 8 18:35:09 2019, gautam, Update, LSC, Locking prep
|
Wed Oct 9 18:15:26 2019, gautam, Update, LSC, PRMI 3f locking
|
Wed Oct 9 22:02:58 2019, gautam, Update, LSC, REFL55 whitening issue
|
Thu Oct 10 01:12:56 2019, gautam, Update, LSC, Locking studies
|
Wed Oct 16 11:42:17 2019, gautam, Update, LSC, Poor separation of PRCL/MICH in 3f signals
|
Thu Oct 17 11:19:28 2019, gautam, Update, LSC, Locking activity last night 
|
Tue Oct 22 00:52:27 2019, gautam, Update, LSC, Locking updates
|
Tue Oct 29 01:24:45 2019, gautam, Update, LSC, More locking updates 
|
Tue Oct 29 15:13:19 2019, gautam, Update, LSC, More locking updates
|
Wed Oct 30 01:27:00 2019, gautam, Update, LSC, More locking updates
|
Wed Oct 30 11:53:41 2019, gautam, Update, LSC, MICH loop shape tuning 
|
Wed Oct 30 17:08:40 2019, gautam, Update, LSC, POP22 investigation
|
Mon Nov 4 15:29:47 2019, gautam, Update, LSC, POP signal path
|
Mon Nov 4 16:06:58 2019, gautam, Update, LSC, POP optical path    
|
Tue Nov 5 11:52:27 2019, gautam, Update, LSC, Locking notes
|
Wed Nov 6 02:08:48 2019, gautam, Update, LSC, Locking updates
|
Wed Nov 6 17:05:45 2019, gautam, Update, LSC, CARM calibration  
|
Wed Nov 6 17:45:34 2019, gautam, Update, LSC, ~
|
Mon Nov 18 21:04:38 2019, gautam, Update, LSC, Locking - some ideas
|
Wed Nov 20 01:07:18 2019, gautam, Update, LSC, Locking - progress
|
Wed Nov 20 12:14:17 2019, gautam, Update, LSC, Locking - progress
|
Wed Nov 20 17:20:24 2019, Yehonathan, Update, LSC, QPD Investigation
|
Wed Nov 20 17:52:00 2019, gautam, Update, LSC, QPD MEDM screen update  
|
Wed Nov 20 21:29:28 2019, gautam, Update, LSC, PRG ~13
|
Fri Nov 22 00:54:14 2019, gautam, Update, LSC, locking notes
|
Mon Nov 25 19:11:22 2019, gautam, Update, LSC, ALS noise re-look
|
Wed Nov 27 23:24:01 2019, gautam, Update, LSC, No shaking but no inspiration either
|
Sun Dec 1 13:38:48 2019, rana, Update, LSC, No shaking but no inspiration either
|
Tue Dec 3 00:03:57 2019, gautam, Update, LSC, Green ALS also shows elevated noise with high arm buildup
|
Wed Dec 4 19:54:27 2019, gautam, Update, LSC, A look to the past
|
Thu Dec 5 01:54:39 2019, gautam, Update, LSC, Partial CM board path engaged
|
Fri Dec 6 00:02:48 2019, gautam, Update, LSC, What is the correct way to set the 3f offsets?
|
Fri Nov 15 12:08:04 2019, gautam, Update, LSC, POPDC whitening board
|
Tue Oct 29 17:40:48 2019, gautam, Update, LSC, More locking updates 
|
|
Message ID: 14997
Entry time: Tue Oct 29 15:13:19 2019
In reply to: 14996
Reply to this: 14999
|
Author: |
gautam |
Type: |
Update |
Category: |
LSC |
Subject: |
More locking updates |
|
|
I looked at some signals for a 10 second period when the PRMI was locked with at some CARM offset, just before the PRMI lost lock, to see if there are any clues. I don't see any obvious signatures in this set of signals - if anything, the PRM is picking up some pitch offset, this is seen both at the Oplev error point and also in the POP QPD spot position. But why should this be happening as I reduce the CARM offset? The arm transmission is only ~5, so it's hard to imagine that the radiation pressure is somehow torquing the PRM. There are no angular feedback loops actuating on the PRM in this state except the local damping and Oplev loops.
The 1f signals are also changing their mean DC offset values, which may be a signature of a changing offset in the 3f MICH and PRCL error points? The MICH error signal is pretty noisy (maybe I can turn on some LPF to clean this up a bit), but I don't see any DC drift in the PRCL control signal. |
|
|