Message ID: 14984
Entry time: Tue Oct 22 15:32:15 2019
In reply to: 14981
Reply to this: 14986
|
Author: |
gautam |
Type: |
Update |
Category: |
ALS |
Subject: |
EX uPDH electronics checkout |
|
|
Summary:
The EX PDH setup had what I thought was insufficient phase and gain margins. So I lowered the gain a little - the price paid was that the suppression of laser frequency noise of the end laser was reduced. I actually think an intermediate gain setting (G=7) can give us ~35 degrees of phase margin, ~10dB gain margin, and lower residual unsuppressed AUX laser noise - to be confirmed by measurement later. See here for the last activity I did - how did the gain get increased? I can't find anything in the elog. |
|
|