40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Jun 3 16:46:31 2019, gautam, Update, IOO, IMC not locking PCdrive_RMS.png
    Reply  Tue Jun 4 10:56:31 2019, gautam, Update, IOO, IMC diagnostics IMCdiag.pdf
       Reply  Thu Jun 6 22:11:53 2019, Koji, Update, IOO, IMC diagnostics 
          Reply  Thu Jun 13 18:01:18 2019, aaron, Update, IOO, IMC diagnostics 
             Reply  Fri Jun 14 13:10:00 2019, aaron, Update, IOO, IMC diagnostics 
                Reply  Mon Jun 17 12:37:16 2019, aaron, Update, IOO, IMC diagnostics 
                   Reply  Tue Jun 18 20:35:07 2019, aaron, Update, IOO, IMC diagnostics 
                      Reply  Wed Jun 19 19:12:51 2019, aaron, Update, IOO, IMC diagnostics PMC_Error_Spectrum.pdfPDH_signal.pdfPDH_signal_full.pdf
                         Reply  Mon Jun 24 15:49:05 2019, aaron, Update, IOO, IMC diagnostics sweep_config_updated.png
                            Reply  Tue Jul 2 19:36:18 2019, aaron, Update, IOO, IMC diagnostics PDH_error.pdfPMC_Error_Spectrum.pdf
          Reply  Fri Jun 21 19:36:26 2019, Koji, Update, IOO, IMC diagnostics IMC_error.pdf
             Reply  Sun Jun 23 08:09:53 2019, gautam, Update, IOO, NPRO diagnostics consolidatedOscopeScreenCaps.pdfRINcomp.pdfPLL_OLTF.pdfPLLnoise.pdf
                Reply  Sun Jun 23 09:36:32 2019, gautam, Update, IOO, IMC is locking normally again PMCdiag.pdf
                   Reply  Sun Jun 23 14:43:14 2019, Koji, Update, IOO, IMC is locking normally again 
                   Reply  Mon Jun 24 08:12:10 2019, gautam, Update, IOO, IMC is locking normally again IMCdutycycle.png
                      Reply  Mon Jun 24 11:48:35 2019, gautam, Update, IOO, IMC in-loop error spectra and OLTF IMCfreqNoise.pdfIMC_OLTF.pdfIMC_CMboard.jpg
Message ID: 14688     Entry time: Sun Jun 23 09:36:32 2019     In reply to: 14687     Reply to this: 14689   14690
Author: gautam 
Type: Update 
Category: IOO 
Subject: IMC is locking normally again 

After typing up the elog, I decided to try locking the IMC again - now it locks again with the "OLD" gain settings. I tested it ~5 times, the autolocker brings the lock back and the PC drive levels are normal. IMC transmission and MC REFL DC light levels in lock are normal. The PC Drive RMS voltage is <1V. What's more, there is no longer any evidence of 60 Hz line harmonics any more in the PMC diagnostics channels. Compare attachment #1 to this elog.

WTF.

I undid the changes Koji made to the autolocker gains, and am trying the old settings again. Let' see how stable or otherwise the config is. I must've jiggled some poor cable connection back into a good spot while working on the PSL table?

Anyway, this helps Kruthi and Milind.

Attachment 1: PMCdiag.pdf  35 kB  Uploaded Sun Jun 23 10:46:02 2019  | Hide | Hide all
PMCdiag.pdf