Entry  Mon Jun 3 16:46:31 2019, gautam, Update, IOO, IMC not locking PCdrive_RMS.png
    Reply  Tue Jun 4 10:56:31 2019, gautam, Update, IOO, IMC diagnostics IMCdiag.pdf
       Reply  Thu Jun 6 22:11:53 2019, Koji, Update, IOO, IMC diagnostics 
          Reply  Thu Jun 13 18:01:18 2019, aaron, Update, IOO, IMC diagnostics 
             Reply  Fri Jun 14 13:10:00 2019, aaron, Update, IOO, IMC diagnostics 
                Reply  Mon Jun 17 12:37:16 2019, aaron, Update, IOO, IMC diagnostics 
                   Reply  Tue Jun 18 20:35:07 2019, aaron, Update, IOO, IMC diagnostics 
                      Reply  Wed Jun 19 19:12:51 2019, aaron, Update, IOO, IMC diagnostics PMC_Error_Spectrum.pdfPDH_signal.pdfPDH_signal_full.pdf
                         Reply  Mon Jun 24 15:49:05 2019, aaron, Update, IOO, IMC diagnostics sweep_config_updated.png
                            Reply  Tue Jul 2 19:36:18 2019, aaron, Update, IOO, IMC diagnostics PDH_error.pdfPMC_Error_Spectrum.pdf
          Reply  Fri Jun 21 19:36:26 2019, Koji, Update, IOO, IMC diagnostics IMC_error.pdf
             Reply  Sun Jun 23 08:09:53 2019, gautam, Update, IOO, NPRO diagnostics consolidatedOscopeScreenCaps.pdfRINcomp.pdfPLL_OLTF.pdfPLLnoise.pdf
                Reply  Sun Jun 23 09:36:32 2019, gautam, Update, IOO, IMC is locking normally again PMCdiag.pdf
                   Reply  Sun Jun 23 14:43:14 2019, Koji, Update, IOO, IMC is locking normally again 
                   Reply  Mon Jun 24 08:12:10 2019, gautam, Update, IOO, IMC is locking normally again IMCdutycycle.png
                      Reply  Mon Jun 24 11:48:35 2019, gautam, Update, IOO, IMC in-loop error spectra and OLTF IMCfreqNoise.pdfIMC_OLTF.pdfIMC_CMboard.jpg
Message ID: 14670     Entry time: Thu Jun 13 18:01:18 2019     In reply to: 14659     Reply to this: 14675
Author: aaron 
Type: Update 
Category: IOO 
Subject: IMC diagnostics 

Continuing this investigation of the IMC, today I am getting familiar with the PMC and FSS. I'd like to measure the frequency noise of the PSL referenced to the PMC.

I checked that the PSL shutter is off, so no light reaches the IMC.

I'm not really sure what I'm looking for on the FSS boxes. I found a few documents to guide:

I ran the FSS autolock script from C1PSL_FSS, nothing obvious changes when I do so. The FSS error signal (which I think is PSL-FSS_MIXERM) is flatlined, and the RC-RF_PD has no LO (PSL-FSS_LOCALC is nan).