Entry  Wed May 15 00:40:19 2019, gautam, Update, SUS, ETMY diagnosis plan ETMY_sensorSpectra.pdf
    Reply  Wed May 15 10:57:57 2019, gautam, Update, SUS, EY chamber opened 
       Reply  Wed May 15 17:46:24 2019, gautam, Update, SUS, ETMY inspection firstContactFiber.JPGETMY_sensorSpectra.pdf
          Reply  Wed May 15 19:36:29 2019, Koji, Update, SUS, ETMY instepction 
             Reply  Thu May 16 13:07:14 2019, gautam, Update, SUS, First contact residue removal etmy_sensors.pdfetmy_BRmode.pdf
                Reply  Thu May 16 23:31:55 2019, gautam, Update, SUS, ETMY suspension characterization 6x
                   Reply  Fri May 17 17:01:08 2019, gautam, Update, SUS, ETMY suspension characterization etmy_oplevs.pdf
          Reply  Mon May 20 11:33:46 2019, gautam, Update, SUS, ITMY inspection 
             Reply  Mon May 20 17:12:57 2019, gautam, Update, SUS, ETMY LL adjustment 
                Reply  Mon May 20 22:06:07 2019, gautam, Update, SUS, ITMY also kicked 
                   Reply  Tue May 21 00:15:21 2019, gautam, Update, SUS, Main objectives of vent achieved (?) etmy_oplevs_20190520.pdfETMY_cplxTF.pdfETMY_diagComp.pdfScreen_Shot_2019-05-21_at_12.37.08_AM.png
                      Reply  Tue May 21 21:33:27 2019, gautam, Update, SUS, ETMY HR face cleaned 
                         Reply  Wed May 22 11:53:50 2019, gautam, Update, SUS, ETMY EQ stops backed out ETMY_eigenmodes.pdf
Message ID: 14612     Entry time: Wed May 15 19:36:29 2019     In reply to: 14611     Reply to this: 14613
Author: Koji 
Type: Update 
Category: SUS 
Subject: ETMY instepction 

A pair of tweezer is OK as long as there is no magnets around. You need to (somewhat) constrain the mirror with the EQ stops so that you can pull the fiber without dragging the mirror.