Entry  Wed May 15 00:40:19 2019, gautam, Update, SUS, ETMY diagnosis plan ETMY_sensorSpectra.pdf
    Reply  Wed May 15 10:57:57 2019, gautam, Update, SUS, EY chamber opened 
       Reply  Wed May 15 17:46:24 2019, gautam, Update, SUS, ETMY inspection firstContactFiber.JPGETMY_sensorSpectra.pdf
          Reply  Wed May 15 19:36:29 2019, Koji, Update, SUS, ETMY instepction 
             Reply  Thu May 16 13:07:14 2019, gautam, Update, SUS, First contact residue removal etmy_sensors.pdfetmy_BRmode.pdf
                Reply  Thu May 16 23:31:55 2019, gautam, Update, SUS, ETMY suspension characterization 6x
                   Reply  Fri May 17 17:01:08 2019, gautam, Update, SUS, ETMY suspension characterization etmy_oplevs.pdf
          Reply  Mon May 20 11:33:46 2019, gautam, Update, SUS, ITMY inspection 
             Reply  Mon May 20 17:12:57 2019, gautam, Update, SUS, ETMY LL adjustment 
                Reply  Mon May 20 22:06:07 2019, gautam, Update, SUS, ITMY also kicked 
                   Reply  Tue May 21 00:15:21 2019, gautam, Update, SUS, Main objectives of vent achieved (?) etmy_oplevs_20190520.pdfETMY_cplxTF.pdfETMY_diagComp.pdfScreen_Shot_2019-05-21_at_12.37.08_AM.png
                      Reply  Tue May 21 21:33:27 2019, gautam, Update, SUS, ETMY HR face cleaned 
                         Reply  Wed May 22 11:53:50 2019, gautam, Update, SUS, ETMY EQ stops backed out ETMY_eigenmodes.pdf
Message ID: 14608     Entry time: Wed May 15 00:40:19 2019     Reply to this: 14610
Author: gautam 
Type: Update 
Category: SUS 
Subject: ETMY diagnosis plan 

I collected some free-swinging data from earlier today evening. There are still only 3 peaks visible in the ASDs, see Attachment #1.

Plan for tomorrow:

TBH, I don't have any clear ideas as to what we are supposed to do to to fix the problem (or even what the problem is). So here is my plan for now:

  1. Take pictures of relative position of magnet and OSEM coil for all five coils
  2. Inspect positions of all EQ stops - back them well out if any look suspiciously close
  3. Inspect suspension wire for any kinks
  4. Inspect position of suspension wire in standoff

I anticipate that these will throw up some more clues 

Attachment 1: ETMY_sensorSpectra.pdf  141 kB  Uploaded Wed May 15 01:47:04 2019  | Hide | Hide all