40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Fri Apr 19 14:06:36 2019, gautam, Update, PSL, Innolight NPRO shutoff
|
Fri Apr 19 20:21:52 2019, gautam, Update, PSL, Innolight NPRO shutoff
|
Wed Apr 24 16:06:44 2019, gautam, Update, PSL, Innolight NPRO shutoff
|
Thu Apr 25 01:03:29 2019, gautam, Update, PSL, MC trans is ~1000 cts (~7%) lower than usual
|
Thu Apr 25 17:31:56 2019, gautam, Update, PSL, Innolight NPRO shutoff 
|
Mon Apr 29 16:25:22 2019, gautam, Update, PSL, PSL turned on again
|
Fri May 3 12:51:58 2019, gautam, Update, PSL, PSL turned on again
|
Mon May 6 10:51:43 2019, gautam, Update, PSL, PSL turned off again
|
Wed May 8 19:04:20 2019, rana, Update, PSL, PSL turned on again
|
Thu May 9 19:50:04 2019, gautam, Update, PSL, PSL turned off again
|
Mon May 13 10:45:38 2019, gautam, Update, PSL, PSL turned ON again
|
Wed May 15 10:56:47 2019, gautam, Update, PSL, PSL turned ON again
|
|
Message ID: 14605
Entry time: Mon May 13 10:45:38 2019
In reply to: 14599
Reply to this: 14609
|
Author: |
gautam |
Type: |
Update |
Category: |
PSL |
Subject: |
PSL turned ON again |
|
|
I used some double-sided tape to attach a San Ace 60 9S0612H4011 to the Innolight controller (Attachment #1). This particular fan is rated to run with up to 13.8V, but I'm using a +15V Sorensen output - at best, this shortens the lifespan of the fan, but I don't have a better solution for now. Then I turned the laser on again (~1040 local time), using the same settings Rana configured earlier in this thread. PMC was locked, and the IMC also could be locked but I closed the shutter for now while the laser frequency/intensity stabilizes after startup. The purpose is to facilitate completion of the pre-vent alignment checklist in prep for the planned vent tomorrow. PMC Trans reports 0.63 after alignment was optimized, which is ~15% lower than in Oct 2016. |
|
|