Entry  Fri Apr 19 14:06:36 2019, gautam, Update, PSL, Innolight NPRO shutoff Screenshot_from_2019-04-19_14-06-11.png
    Reply  Fri Apr 19 20:21:52 2019, gautam, Update, PSL, Innolight NPRO shutoff Screenshot_from_2019-04-19_20-27-04.png
       Reply  Wed Apr 24 16:06:44 2019, gautam, Update, PSL, Innolight NPRO shutoff PMCrefl.png
          Reply  Thu Apr 25 01:03:29 2019, gautam, Update, PSL, MC trans is ~1000 cts (~7%) lower than usual 
          Reply  Thu Apr 25 17:31:56 2019, gautam, Update, PSL, Innolight NPRO shutoff EdwinShutoff20190425.pngEdwinShutdown_zoomOut.png
             Reply  Mon Apr 29 16:25:22 2019, gautam, Update, PSL, PSL turned on again 
                Reply  Fri May 3 12:51:58 2019, gautam, Update, PSL, PSL turned on again 
                   Reply  Mon May 6 10:51:43 2019, gautam, Update, PSL, PSL turned off again NPROshutoff.png
                   Reply  Wed May 8 19:04:20 2019, rana, Update, PSL, PSL turned on again 
                      Reply  Thu May 9 19:50:04 2019, gautam, Update, PSL, PSL turned off again Screenshot_from_2019-05-09_19-49-29.png
                         Reply  Mon May 13 10:45:38 2019, gautam, Update, PSL, PSL turned ON again IMG_7431.JPG
                            Reply  Wed May 15 10:56:47 2019, gautam, Update, PSL, PSL turned ON again 
Message ID: 14577     Entry time: Thu Apr 25 17:31:56 2019     In reply to: 14566     Reply to this: 14583
Author: gautam 
Type: Update 
Category: PSL 
Subject: Innolight NPRO shutoff 

NPRO shutoff at ~1517  local time today afternoon. Again, not many clues from the NPRO diagnostics channel, but to my eye, the D1_POW channel shows the first variation from the "steady state", followed by the other channels. This is ~0.1 sec before the other channels register some change, so I don't know how much we can trust the synchronizaiton of the EPICS data streams. I won't turn it on again for now. I did check that the little fan on the back of the NPRO controller is still rotating.

gautam 10am 4/29: I also added a longer term trend of these diagnostic channels, no clear trends suggesting a fault are visible. The y-axis units for all plots are in Volts, and the data is sampled at 16 Hz.

Quote:

Now we wait and watch I guess.

Attachment 1: EdwinShutoff20190425.png  91 kB  Uploaded Thu Apr 25 18:32:34 2019  | Hide | Hide all
EdwinShutoff20190425.png
Attachment 2: EdwinShutdown_zoomOut.png  76 kB  Uploaded Mon Apr 29 11:23:46 2019  | Hide | Hide all
EdwinShutdown_zoomOut.png