Entry  Thu Apr 25 10:32:39 2019, Jon, Update, VAC, Vac interlocks updated 
    Reply  Thu Apr 25 11:27:11 2019, gautam, Update, VAC, PSL shutter re-opened 
Message ID: 14575     Entry time: Thu Apr 25 11:27:11 2019     In reply to: 14574
Author: gautam 
Type: Update 
Category: VAC 
Subject: PSL shutter re-opened 

This activity seems to have closed the PSL shutter (actually I'm not sure why that happened - the interlock should only trip if P1a exceeds 3 mtorr, and looking at the time series for the last 2 hours, it did not ever exceed this threshold). I saw no reason for it to remain closed so I re-opened it just now.

I vote for not remotely rebooting any of the vacuum / PSL subsystems. In the event of something going catastrophically wrong, someone should be on hand to take action in the lab.