Entry  Fri Apr 19 14:06:36 2019, gautam, Update, PSL, Innolight NPRO shutoff Screenshot_from_2019-04-19_14-06-11.png
    Reply  Fri Apr 19 20:21:52 2019, gautam, Update, PSL, Innolight NPRO shutoff Screenshot_from_2019-04-19_20-27-04.png
       Reply  Wed Apr 24 16:06:44 2019, gautam, Update, PSL, Innolight NPRO shutoff PMCrefl.png
          Reply  Thu Apr 25 01:03:29 2019, gautam, Update, PSL, MC trans is ~1000 cts (~7%) lower than usual 
          Reply  Thu Apr 25 17:31:56 2019, gautam, Update, PSL, Innolight NPRO shutoff EdwinShutoff20190425.pngEdwinShutdown_zoomOut.png
             Reply  Mon Apr 29 16:25:22 2019, gautam, Update, PSL, PSL turned on again 
                Reply  Fri May 3 12:51:58 2019, gautam, Update, PSL, PSL turned on again 
                   Reply  Mon May 6 10:51:43 2019, gautam, Update, PSL, PSL turned off again NPROshutoff.png
                   Reply  Wed May 8 19:04:20 2019, rana, Update, PSL, PSL turned on again 
                      Reply  Thu May 9 19:50:04 2019, gautam, Update, PSL, PSL turned off again Screenshot_from_2019-05-09_19-49-29.png
                         Reply  Mon May 13 10:45:38 2019, gautam, Update, PSL, PSL turned ON again IMG_7431.JPG
                            Reply  Wed May 15 10:56:47 2019, gautam, Update, PSL, PSL turned ON again 
Message ID: 14570     Entry time: Thu Apr 25 01:03:29 2019     In reply to: 14566
Author: gautam 
Type: Update 
Category: PSL 
Subject: MC trans is ~1000 cts (~7%) lower than usual 

When dialing up the current, I went up to 2.01 A on the front panel display, which is what I remember it being. The label on the controller is from when the laser was still putting out 2W, and says the pump current should be 2.1 A. Anyhow, the MC transmission is ~7% lower now (14500 cts compared to the usual 15000-15500 cts), even after tweaking the PMC alignment to minimize PMC REFL. Potentially there is less power coming out of the NPRO. I will measure it at the window tomorrow with a power meter.