40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sat Apr 6 02:07:40 2019, Anjali, Update, Frequency noise measurement, Frequency noise measurement of 1 micron source 
    Reply  Wed Apr 10 00:33:09 2019, Anjali, Update, Frequency noise measurement, Frequency noise measurement of 1 micron source Modulation_frequency_100Hz.jpgModulation_frequency_30Hz.jpgModified_setup.JPG
       Reply  Wed Apr 10 16:58:54 2019, rana, Update, IOO, fiber MZ for NPRO freq noise measurements 
          Reply  Wed Apr 10 22:59:22 2019, gautam, Update, IOO, Spooled fiber 
             Reply  Thu Apr 11 09:05:06 2019, Anjali, Update, IOO, Spooled fiber Frequecy_modulation_32_Hz.pdfFrequecy_modulation_140_Hz.pdfFrequecy_modulation_300_Hz.pdfWithout_modulation.pdfNew_fiber_spool.JPG
       Reply  Fri Apr 12 01:22:27 2019, Anjali, Update, Frequency noise measurement, Frequency noise measurement of 1 micron source 
          Reply  Mon Apr 15 22:39:10 2019, gautam, Update, Frequency noise measurement, Alternate setup with PSL pickoff 
             Reply  Wed Apr 17 00:43:38 2019, gautam, Update, Frequency noise measurement, MZ interferometer ---> DAQ 
                Reply  Thu Apr 25 03:32:25 2019, Anjali, Update, Frequency noise measurement, MZ interferometer ---> DAQ Time_domain_output.pdfFrequency_noise.pdfschematic_heterodyne_setup.pngNoise_budget_1_micron_in_Hz_per_rtHz.pdf
                   Reply  Thu Apr 25 10:25:19 2019, gautam, Update, Frequency noise measurement, Homodyne v Heterodyne 
                      Reply  Thu Apr 25 15:47:54 2019, Anjali, Update, Frequency noise measurement, Homodyne v Heterodyne Experimental_setup.JPG
Message ID: 14534     Entry time: Thu Apr 11 09:05:06 2019     In reply to: 14531
Author: Anjali 
Type: Update 
Category: IOO 
Subject: Spooled fiber 
  • Attchment #1,2,3 and 4 shows the results with frequency modulation of 32 Hz, 140 Hz , 300 Hz and without frequency modulation. I am trying to understand these results better.
  • A lot of fringing is there even when no modulation is applied. We hope to improve this by spooling the fiber and then encasing it in a box. 
  • As mentioned by Gautam, we have got a 50 m spooled fiber. Attachment #5 shows the photo of the same
Quote:

Steve had showed me some stock of long fibers a while back - they are from Oz Optics, and are 50m long, and are already spooled - so barring objections, we will try the MZ setup with the spooled fiber and see if there is any improvement in the fringing rate of the MZ. Then we can evaluate what additional stabilization of the fiber length is required. Anjali will upload a photo of the spooled fiber.

Attachment 1: Frequecy_modulation_32_Hz.pdf  8 kB  Uploaded Thu Apr 11 10:06:17 2019  | Hide | Hide all
Frequecy_modulation_32_Hz.pdf
Attachment 2: Frequecy_modulation_140_Hz.pdf  60 kB  Uploaded Thu Apr 11 10:06:32 2019  | Hide | Hide all
Frequecy_modulation_140_Hz.pdf
Attachment 3: Frequecy_modulation_300_Hz.pdf  62 kB  Uploaded Thu Apr 11 10:06:49 2019  | Hide | Hide all
Frequecy_modulation_300_Hz.pdf
Attachment 4: Without_modulation.pdf  78 kB  Uploaded Thu Apr 11 10:14:09 2019  | Hide | Hide all
Without_modulation.pdf
Attachment 5: New_fiber_spool.JPG  4.296 MB  Uploaded Thu Apr 11 10:14:37 2019  | Hide | Hide all
New_fiber_spool.JPG
ELOG V3.1.3-