Yoichi, Peter
We suspected that maybe the 3.8kHz peak is the DARM RSE somehow coupled to the CARM.
So we added an offset to the SRC error signal to see if the peak moves by changing the offset.
It didn't (at least by changing the SRC offset by +/-1000).
(I had a nice plot showing this, but dtt corrupted the data when I saved it. So no plot attached.)
I also played with the PRC, DARM offsets which did not have any effect on the peak.
The only thing, I could find so far, having some effect on the peak is the arm power. As the arm power is increased, the peak height goes up and the frequency shifts slightly towards lower frequencies. |