40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Feb 20 23:55:51 2019, gautam, Update, ALS, ALS delay line electronics delayLineDemod.pdf.zip
    Reply  Thu Mar 7 01:06:38 2019, gautam, Update, ALS, ALS delay line electronics 
       Reply  Tue Mar 12 22:51:25 2019, gautam, Update, ALS, ALS delay line electronics 
       Reply  Wed Mar 13 01:27:30 2019, gautam, Update, ALS, ALS delay line electronics DFDcal.pdfDFDnoise.pdf
          Reply  Sun Mar 17 00:42:20 2019, gautam, Update, ALS, NF1611 cannot be shot-noise limited? PDnoise.pdf
             Reply  Thu Mar 28 19:40:02 2019, gautam, Update, ALS, BeatMouth with NF1611s assembled IMG_7381.JPGIMG_7382.JPGrelTF_schem.pdfdarkNoise.pdf
                Reply  Fri Mar 29 21:00:06 2019, gautam, Update, ALS, BeatMouth with NF1611s installed IMG_7384.JPGIMG_7385.JPG
                   Reply  Sun Mar 31 15:05:53 2019, gautam, Update, ALS, Fiber beam-splitters not PM IMG_7384.JPG
                      Reply  Wed Apr 3 09:04:01 2019, gautam, Update, ALS, Note about new fiber couplers 
                         Reply  Wed Apr 3 12:32:33 2019, Koji, Update, ALS, Note about new fiber couplers 
                      Reply  Fri Apr 5 00:33:58 2019, gautam, Update, ALS, Promising IR ALS noise IMG_7388.JPGfreeSpace_IR_beat.pdf
                         Reply  Fri Apr 5 11:49:30 2019, gautam, Update, ALS, PSL + X green beat recovery 
                         Reply  Mon Apr 8 00:04:08 2019, gautam, Update, ALS, IR ALS noise budget DFDnoise.pdf
                            Reply  Mon Apr 8 18:28:25 2019, gautam, Update, ALS, EX Green PDH checkout IMG_7393.JPGgreenModeMatch.pdf
                               Reply  Mon Apr 8 23:52:09 2019, gautam, Update, ALS, EX Green PDH error monitor calibrated errMonCalib.pdferrMon.pngOLTF.pdfEX_frequencyNoises.pdf
                                  Reply  Tue Apr 9 00:16:22 2019, rana, Update, ALS, EX Green PDH error monitor calibrated 
                                     Reply  Tue Apr 9 00:18:19 2019, gautam, Update, ALS, EX Green PDH error monitor calibrated 
                                  Reply  Tue Apr 9 18:44:00 2019, gautam, Update, ALS, EX Green PDH discriminant measurement cavityFlashes.pdfmodelPDH.pdf
                            Reply  Tue Apr 9 19:07:12 2019, gautam, Update, ALS, IR ALS noise budget ALS_noiseBudget.pdfALS_noiseBudget.pdf
                               Reply  Thu Apr 11 01:10:05 2019, gautam, Update, ALS, Large 2kHz peak (and harmonics) in ALS X EX_PDH_2kNoise.pdf
                                  Reply  Wed Apr 17 00:50:17 2019, gautam, Update, ALS, Large 2kHz peak (and harmonics) in ALS X no more EX_PDHnoise.pdf
                                     Reply  Wed Apr 17 11:01:49 2019, gautam, Update, ALS, Large 2kHz peak (and harmonics) in ALS X no more EX_PDH_specGram.pdf
                      Reply  Wed May 29 18:13:25 2019, gautam, Update, ALS, Fiber beam-splitters are now PM 
                         Reply  Fri May 31 15:55:16 2019, gautam, Update, ALS, PSL + X beat restored 
Message ID: 14480     Entry time: Sun Mar 17 00:42:20 2019     In reply to: 14478     Reply to this: 14498
Author: gautam 
Type: Update 
Category: ALS 
Subject: NF1611 cannot be shot-noise limited? 

Summary:

Per the manual (pg12) of the NF 1611 photodiode, the "Input Noise Current" is 16 pA/rtHz. It also specifies that for "Linear Operation", the max input power is 1 mW, which at 1um corresponds to a current shot noise of ~14 pA/rtHz. Therefore,

  1. This photodiode cannot be shot-noise limited if we also want to stay in the spec-ed linear regime.
  2. We don't need to worry so much about the noise figure of the RF amplifier that follows the photodiode. In fact, I think we can use a higher gain RF amplifier with a slightly worse noise figure (e.g. ZHL-3A) as we will benefit from having a larger frequency discriminant with more RF power reaching the delay line.

Details:

Attachment #1: Here, I plot the expected voltage noise due to shot noise of the incident light, assuming 0.75 A/W for InGaAs and 700V/A transimpedance gain. 

  • For convenience, I've calibrated on the twin axes the current shot noise (X) and equivalent amplifier noise figure at a given voltage noise, assuming a 50 ohm system (Y).
  • The 16 pA/rtHz input current noise exceeds the shot noise contribution for powers as high as 1 mW.
  • Even at 0.5 mW power on the PD, we can use the ZHL-3A rather than the Teledyne:
    • This calculation was motivated by some suspicious features in the Teledyne amplifier gain, I will write a separate elog about that. 
    • For the light levels we have, I expect ~3dBm RF signal from the photodiode. With the 24dB of gain from the ZHL-3A, the signal becomes 27dBm, which is smaller (but close to) the spec-ed max output of the ZHL-3A, which is 29.5 dBm. Is this too close to the edge?
    • I will measure the gain/noise of the ZHL-3A to get a better answer to these questions.
  • If in the future we get a better photodiode setup that reaches sub-1nV/rtHz (dark/electronics) voltage noise, we may have to re-evaluate what is an appropriate RF amplifier.