40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Wed Feb 20 23:55:51 2019, gautam, Update, ALS, ALS delay line electronics
|
Thu Mar 7 01:06:38 2019, gautam, Update, ALS, ALS delay line electronics
|
Tue Mar 12 22:51:25 2019, gautam, Update, ALS, ALS delay line electronics
|
Wed Mar 13 01:27:30 2019, gautam, Update, ALS, ALS delay line electronics 
|
Sun Mar 17 00:42:20 2019, gautam, Update, ALS, NF1611 cannot be shot-noise limited?
|
Thu Mar 28 19:40:02 2019, gautam, Update, ALS, BeatMouth with NF1611s assembled   
|
Fri Mar 29 21:00:06 2019, gautam, Update, ALS, BeatMouth with NF1611s installed 
|
Sun Mar 31 15:05:53 2019, gautam, Update, ALS, Fiber beam-splitters not PM
|
Wed Apr 3 09:04:01 2019, gautam, Update, ALS, Note about new fiber couplers
|
Wed Apr 3 12:32:33 2019, Koji, Update, ALS, Note about new fiber couplers
|
Fri Apr 5 00:33:58 2019, gautam, Update, ALS, Promising IR ALS noise 
|
Fri Apr 5 11:49:30 2019, gautam, Update, ALS, PSL + X green beat recovery
|
Mon Apr 8 00:04:08 2019, gautam, Update, ALS, IR ALS noise budget
|
Mon Apr 8 18:28:25 2019, gautam, Update, ALS, EX Green PDH checkout 
|
Mon Apr 8 23:52:09 2019, gautam, Update, ALS, EX Green PDH error monitor calibrated   
|
Tue Apr 9 00:16:22 2019, rana, Update, ALS, EX Green PDH error monitor calibrated
|
Tue Apr 9 00:18:19 2019, gautam, Update, ALS, EX Green PDH error monitor calibrated
|
Tue Apr 9 18:44:00 2019, gautam, Update, ALS, EX Green PDH discriminant measurement 
|
Tue Apr 9 19:07:12 2019, gautam, Update, ALS, IR ALS noise budget 
|
Thu Apr 11 01:10:05 2019, gautam, Update, ALS, Large 2kHz peak (and harmonics) in ALS X
|
Wed Apr 17 00:50:17 2019, gautam, Update, ALS, Large 2kHz peak (and harmonics) in ALS X no more
|
Wed Apr 17 11:01:49 2019, gautam, Update, ALS, Large 2kHz peak (and harmonics) in ALS X no more
|
Wed May 29 18:13:25 2019, gautam, Update, ALS, Fiber beam-splitters are now PM
|
Fri May 31 15:55:16 2019, gautam, Update, ALS, PSL + X beat restored
|
|
Message ID: 14480
Entry time: Sun Mar 17 00:42:20 2019
In reply to: 14478
Reply to this: 14498
|
Author: |
gautam |
Type: |
Update |
Category: |
ALS |
Subject: |
NF1611 cannot be shot-noise limited? |
|
|
Summary:
Per the manual (pg12) of the NF 1611 photodiode, the "Input Noise Current" is 16 pA/rtHz. It also specifies that for "Linear Operation", the max input power is 1 mW, which at 1um corresponds to a current shot noise of ~14 pA/rtHz. Therefore,
- This photodiode cannot be shot-noise limited if we also want to stay in the spec-ed linear regime.
- We don't need to worry so much about the noise figure of the RF amplifier that follows the photodiode. In fact, I think we can use a higher gain RF amplifier with a slightly worse noise figure (e.g. ZHL-3A) as we will benefit from having a larger frequency discriminant with more RF power reaching the delay line.
Details:
Attachment #1: Here, I plot the expected voltage noise due to shot noise of the incident light, assuming 0.75 A/W for InGaAs and 700V/A transimpedance gain.
- For convenience, I've calibrated on the twin axes the current shot noise (X) and equivalent amplifier noise figure at a given voltage noise, assuming a 50 ohm system (Y).
- The 16 pA/rtHz input current noise exceeds the shot noise contribution for powers as high as 1 mW.
- Even at 0.5 mW power on the PD, we can use the ZHL-3A rather than the Teledyne:
- This calculation was motivated by some suspicious features in the Teledyne amplifier gain, I will write a separate elog about that.
- For the light levels we have, I expect ~3dBm RF signal from the photodiode. With the 24dB of gain from the ZHL-3A, the signal becomes 27dBm, which is smaller (but close to) the spec-ed max output of the ZHL-3A, which is 29.5 dBm. Is this too close to the edge?
- I will measure the gain/noise of the ZHL-3A to get a better answer to these questions.
- If in the future we get a better photodiode setup that reaches sub-1nV/rtHz (dark/electronics) voltage noise, we may have to re-evaluate what is an appropriate RF amplifier.
|
|
|