Wed Jan 9 20:01:25 2019, Jon, Update, VAC, Second pumpdown completed 
|
Thu Jan 10 10:23:46 2019, gautam, Update, VAC, overnight leak rate
|
Sat Feb 2 21:53:24 2019, Koji, Update, VAC, overnight leak rate
|
Sun Feb 3 15:15:21 2019, gautam, Update, VAC, overnight leak rate 
|
Sun Feb 3 20:52:34 2019, Koji, Update, VAC, overnight leak rate
|
Mon Feb 4 12:23:24 2019, gautam, Update, VAC, pumpdown 83 - leak tests
|
Tue Feb 5 10:11:30 2019, gautam, Update, VAC, leak tests complete, pumpdown 83 resumed 
|
Tue Feb 5 19:30:14 2019, gautam, Update, VAC, Main volume at 20 uTorr
|
Thu Feb 7 13:55:25 2019, gautam, Update, VAC, RGA turned on
|
Thu Feb 7 19:28:46 2019, gautam, Update, VAC, IFO recovery
|
|
Message ID: 14429
Entry time: Sat Feb 2 21:53:24 2019
In reply to: 14394
Reply to this: 14430
|
Author: |
Koji |
Type: |
Update |
Category: |
VAC |
Subject: |
overnight leak rate |
|
|
The pressure of the main volume increased from ~1mtorr to 50mtorr for the past 24 hours (86ksec). This rate is about x1000 of the reported number on Jan 10. Do we suspect vacuum leak?
Quote: |
Overnight, the pressure increased from 247 uTorr to 264 uTorr over a period of 30000 seconds. Assuming an IFO volume of 33,000 liters, this corresponds to an average leak rate of ~20 uTorr L / s.
|
|
|
|