Tue Jan 15 10:52:38 2019, gautam, Update, SUS, EY door opened
|
Tue Jan 15 15:49:47 2019, gautam, Update, SUS, EY door opened
|
Sat Jan 19 15:33:18 2019, gautam, Update, SUS, ETMY OSEMs faulty 
|
Tue Jan 22 20:36:53 2019, gautam, Update, SUS, ETMY OSEMs faulty
|
Wed Jan 23 12:39:18 2019, gautam, Update, SUS, EY chamber work 6x
|
Wed Jan 23 23:12:44 2019, gautam, Update, SUS, Prep for FC cleaning
|
Thu Jan 24 15:32:31 2019, gautam, Update, SUS, Y arm cavity side first contact applied 
|
Tue Jan 29 22:12:40 2019, gautam, Update, SUS, Alignment prep
|
Wed Jan 30 11:54:24 2019, gautam, Update, SUS, More alignment prep 
|
Wed Jan 30 19:25:40 2019, gautam, Update, SUS, Xarm cavity alignment 
|
Fri Feb 1 01:24:06 2019, gautam, Update, SUS, Almost ready for pumpdown tomorrow
|
Fri Feb 1 13:16:50 2019, gautam, Update, SUS, Pumpdown 83 underway
|
Fri Feb 1 21:52:57 2019, gautam, Update, SUS, Pumpdown 83 underway
|
Fri Feb 1 14:44:14 2019, gautam, Update, SUS, Y arm FC cleaning and reinstall  
|
Wed Jan 16 16:25:25 2019, gautam, Update, SUS, Yarm locked
|
Fri Jan 18 21:34:18 2019, gautam, Update, SUS, Unused optic on EY table
|
Sat Jan 19 05:07:45 2019, Koji, Update, SUS, Unused optic on EY table
|
|
Message ID: 14424
Entry time: Wed Jan 30 19:25:40 2019
In reply to: 14423
Reply to this: 14425
|
Author: |
gautam |
Type: |
Update |
Category: |
SUS |
Subject: |
Xarm cavity alignment |
|
|
Squishing cables at the ITMX satellite box seems to have fixed the wandering ITM that I observed yesterday - the sooner we are rid of these evil connectors the better.
I had changed the input pointing of the green injection from EX to mark a "good" alignment of the cavity axis, so I used the green beam to try and recover the X arm alignment. After some tweaking of the ITM and ETM angle bias voltages, I was able to get good GTRX values [Attachment #1], and also see clear evidence of (admittedly weak) IR resonances in TRX [Attachment #2]. I can't see the reflection from ITMX on the AS camera, but I suspect this is because the ITMY cage is in the way. This will likely have to be redone tomorrow after setting the input pointing for the Y arm cavity axis, but hopefully things will converge faster and we can close up sooner. Closing the PSL shutter for now...