Entry  Fri Mar 27 02:40:06 2009, pete, Summary, IOO, MC glitch investigation MC1_Drift.png
    Reply  Fri Mar 27 15:05:42 2009, Yoichi, Update, IOO, MC glitch investigation MC1_Drift.pdfMC2_Drift.pdf
       Reply  Fri Mar 27 17:52:16 2009, Yoichi, Update, IOO, MC glitch investigation 
Message ID: 1437     Entry time: Fri Mar 27 15:05:42 2009     In reply to: 1435     Reply to this: 1438
Author: Yoichi 
Type: Update 
Category: IOO 
Subject: MC glitch investigation 
Attached plots are the result of the MC1 trend measurement.
See the attachment #1. The first two plots show the drift of the MC1 alignment as seen by the OSEMs. It is terrible. Other MC mirrors also drifted but the scale is smaller than the MC1.
From the VMon channels, you can see that the control voltages were quiet.
 
The monitor channels we added were:
 
MC_TMP1 = UL coil bias. Input to the coil driver board.
MC_DRUM1 = UL coil bias. Output of the current buffer.
OSA_APTEMP = LR coil bias. Input to the coil driver board.
OSA_SPTEMP = LR coil bias. Output of the current buffer.
 
The bias voltages show no drift except for a glitch around 7AM. This glitch did not show up in the SPTEMP channel (LR coil bias output). This was because the probe was connected to the coil side of the output resistor by mistake.
 
The second attachment shows a zoomed plot of MC1 OSEM signals along with the bias monitor channels (signals were appropriately scaled so that they all fit in +/-1).
There is no correlation between the OSEM signals and the bias voltages.
 
Since we were only monitoring UL and LR coils, I changed the monitor points as follows.
 
MC_TMP1 = LL coil bias. Output of the current buffer.
MC_DRUM1 = UL coil bias. Output of the current buffer.
OSA_APTEMP = UR coil bias. Output of the current buffer.
OSA_SPTEMP = LR coil bias. Output of the current buffer.
 
I will leave the MC unlocked for a while.

Quote:

Yoichi, Pete

The MC loses lock due to glitches in the MC1 coils. 
We do not know which coil for sure, and we do not know if it is a problem going into the board, or a problem on the board. 
We suspect either the UL or LR coil bias circuits (Pete would bet on UL).  If you look at the bottom 4 plots in the attached file, you can see a relatively large 3 minute dip in the UL OSEM output, with a corresponding bump in the LR (and smaller dips in the other diagonal).  
These bumps do not show up in the VMONS which is why we are suspicious of the bias.
To test we are monitoring 4 points in test channels, for UL and UR, both going into the bias driver circuit, and coming out of the current buffer before going into the coils. 
 

We ran cable from the suspension rack to the IOO rack to record the signals with DAQ channels.

The test channels:

UL coil      C1:IOO-MC_DRUM1  (Caryn was using, we will replace when we are done)

UL input   C1:IOO-MC_TMP1 (Caryn was using, we will replace when we are done)

LR coil      C1:PEM-OSA_SPTEMP

LR input   C1:PEM-OSA_APTEMP

We will leave these overnight; we intend to remove them tomorrow or Monday.

We closed the PSL shutter and killed the MC autolocker.

 

Attachment 1: MC1_Drift.pdf  31 kB  | Hide | Hide all | Show all