Summary:
- More DRMI characterization was done.
- I was working on trying to improve the stability of the DRMI locks as this is necessary for any serious characterization.
- Today I revived the PRC angular feedforward - this was a gamechanger, the DRMI locks were much more stable. It's probably worth spending some time improving the POP LSC/ASC sensing optics/electronics looking towards the full IFO locking.
- Quantitatively, the angular fluctuations as witnessed by the POP QPD is lowered by ~2x with the FF on compared to off
[Attachment #1, references are with FF off, live traces are with FF on].
- The first DRMI lock I got is already running 15 mins, looking stable.
- Update: Out of the ~1 hour i've tried DRMI locking tonight, >50 mins locked!
- I think the filters can be retrained and this performance improved, something to work on while we are vented.
- Ran sensing lines, measured loop TFs, analysis tomorrow, but I think the phasing of the 1f PDs is now okay.
- MICH in AS55 Q, demod phase = -92deg, +6dB wht gain.
- PRCL in REFL11 I, demod phase = +18 deg, +18dB wht gain.
- SRCL in REFL55 I, demod phase = -175 deg, +18dB wht gain.
- Also repeated the line in SRCL-->witness in MICH test.
- At least 10 minutes of data available, but I'm still collecting since the lock is holding.
- This time I drove the line at ~124 Hz with awggui, since this is more a regime where we are sensing noise dominated.
Prep for this work:
- Reboots of c1psl, c1iool0, c1susaux.
- Removed AS port PD loss measurement PD.
- Initial alignment procedure as usual: single arms --> PRMI locked on carrier --> DRMI
I was trying to get some pics of the optics as a zeroth-level reference for the pre-vent loss with the single arms locked, but since our SL7 upgrade, the sensoray won't work anymore . I'll try fixing this during the daytime. |