40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Wed Sep 6 01:09:11 2017, johannes, Update, Computer Scripts / Programs, New set of loss measurements 
|
Mon Sep 11 12:56:40 2017, johannes, Update, Computer Scripts / Programs, lossmap attempts
|
Mon Sep 11 16:46:09 2017, Steve, Update, PEM, earthquakes
|
Tue Sep 19 14:00:41 2017, Steve, Update, PEM, earthquakes
|
Fri Oct 27 09:34:20 2017, Steve, Update, PEM, earthquakes
|
Fri Nov 10 13:46:26 2017, Steve, Update, PEM, M3.1 local earthquake
|
Mon Apr 9 08:39:39 2018, Steve, Update, PEM, M5.3 eq Souther CA  
|
Thu Jan 25 08:27:25 2018, Steve, Update, PEM, earthquakes  
|
Mon Sep 25 17:56:33 2017, johannes, Update, Computer Scripts / Programs, transmitted power during lossmap
|
Thu Oct 11 13:40:51 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Fri Oct 12 12:29:34 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Fri Oct 12 20:20:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Sat Oct 13 20:11:10 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Mon Oct 15 10:32:13 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
|
Tue Oct 16 00:44:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Fri Nov 2 19:25:16 2018, gautam, Update, Computer Scripts / Programs, loss measurements
|
Mon Oct 15 12:52:54 2018, yuki, Update, Computer Scripts / Programs, additional comments
|
Fri Nov 2 16:42:31 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents
|
Mon Nov 5 13:52:18 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents
|
Tue Nov 6 10:19:26 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents
|
Tue Nov 6 15:23:48 2018, gautam, Update, IOO, IMC problematic
|
Tue Nov 6 19:02:35 2018, aaron, Update, IOO, IMC problematic
|
Sat Nov 10 17:40:00 2018, aaron, Update, IOO, IMC problematic
|
Mon Nov 12 13:53:20 2018, rana, Update, IOO, loss measurement: oscope vs CDS DAQ
|
Thu Nov 15 10:21:07 2018, aaron, Update, IOO, IMC problematic 
|
Fri Nov 16 10:53:07 2018, aaron, Update, IOO, IMC problematic
|
Sat Nov 17 18:59:01 2018, aaron, Update, IOO, IMC problematic
|
Wed Nov 7 05:16:16 2018, yuki, Update, Computer Scripts / Programs, arm loss measuremenents
|
|
Message ID: 14289
Entry time: Sat Nov 10 17:40:00 2018
In reply to: 14277
Reply to this: 14290
14297
|
Author: |
aaron |
Type: |
Update |
Category: |
IOO |
Subject: |
IMC problematic |
|
|
Gautam was doing some DRMI locking, so I replaced the photodiode at the AS port to begin loss measurements again.
I increased the resolution on the scope by selecting Average (512) mode. I was a bit confused by this, since Yuki was correct that I had only 4 digits recorded over ethernet, which made me think this was an i/o setting. However the sample acquisition setting was the only thing I could find on the tektronix scope or in its manual about improving vertical resolution. This didn't change the saved file, but I found the more extensive programming manual for the scope, which confirms that using average mode does increase the resolution... from 9 to 14 bits! I'm not even getting that many.
There's another setting for DATa:WIDth, that is the number of bytes per data point transferred from the scope.
I tried using the *.25 scope instead, no better results. Changing the vertical resolution directly doesn't change this either. I've also tried changing most of the ethernet settings. I don't think it's something on the scripts side, because I'm using the same scripts that apparently generated the most recent of Johannes' and Yuki's files; I did look through for eg tds3014b.py, and didn't see the resolution explicitly set. Indeed, I get 7 bits of resolution as that function specifies, but most of them aren't filled by the scope. This makes me think the problem is on the scope settings. |