40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Wed Sep 6 01:09:11 2017, johannes, Update, Computer Scripts / Programs, New set of loss measurements 
|
Mon Sep 11 12:56:40 2017, johannes, Update, Computer Scripts / Programs, lossmap attempts
|
Mon Sep 11 16:46:09 2017, Steve, Update, PEM, earthquakes
|
Tue Sep 19 14:00:41 2017, Steve, Update, PEM, earthquakes
|
Fri Oct 27 09:34:20 2017, Steve, Update, PEM, earthquakes
|
Fri Nov 10 13:46:26 2017, Steve, Update, PEM, M3.1 local earthquake
|
Mon Apr 9 08:39:39 2018, Steve, Update, PEM, M5.3 eq Souther CA  
|
Thu Jan 25 08:27:25 2018, Steve, Update, PEM, earthquakes  
|
Mon Sep 25 17:56:33 2017, johannes, Update, Computer Scripts / Programs, transmitted power during lossmap
|
Thu Oct 11 13:40:51 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Fri Oct 12 12:29:34 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Fri Oct 12 20:20:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Sat Oct 13 20:11:10 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Mon Oct 15 10:32:13 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
|
Tue Oct 16 00:44:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements
|
Fri Nov 2 19:25:16 2018, gautam, Update, Computer Scripts / Programs, loss measurements
|
Mon Oct 15 12:52:54 2018, yuki, Update, Computer Scripts / Programs, additional comments
|
Fri Nov 2 16:42:31 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents
|
Mon Nov 5 13:52:18 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents
|
Tue Nov 6 10:19:26 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents
|
Tue Nov 6 15:23:48 2018, gautam, Update, IOO, IMC problematic
|
Tue Nov 6 19:02:35 2018, aaron, Update, IOO, IMC problematic
|
Sat Nov 10 17:40:00 2018, aaron, Update, IOO, IMC problematic
|
Mon Nov 12 13:53:20 2018, rana, Update, IOO, loss measurement: oscope vs CDS DAQ
|
Thu Nov 15 10:21:07 2018, aaron, Update, IOO, IMC problematic 
|
Fri Nov 16 10:53:07 2018, aaron, Update, IOO, IMC problematic
|
Sat Nov 17 18:59:01 2018, aaron, Update, IOO, IMC problematic
|
Wed Nov 7 05:16:16 2018, yuki, Update, Computer Scripts / Programs, arm loss measuremenents
|
|
Message ID: 14268
Entry time: Fri Nov 2 16:42:31 2018
In reply to: 14255
Reply to this: 14270
|
Author: |
aaron |
Type: |
Update |
Category: |
Computer Scripts / Programs |
Subject: |
arm loss measuremenents |
|
|
I'm continuing the arm loss measurements Yuki was making. I'm first familiarizing myself with the procedures for the measurement Johannes describes.
I'm not very familiar with the medm screens, so I'm just kind of poking around and checking with Gautam. I do the following:
- Turned Xarm ASS dither on, then off.
- Turned X and Y ALS on, then off shortly after
- Realizing I needed some guidance, I found this page on lock acquisition on the wiki
- Gautam showed me how to align/lock the IFO so I could take some notes, and we locked the Y arm, misaligned X.
- I put the PD back in the AS beam path to get the ASDC signal, and approximately centered the beam. This PD is on channel 1 of the scope, which is at 192.168.113.24.
- I centered the beam onto the MC2 PD that Yuki had installed. This PD is on channel 2 of the scope.
- Both scope channels are set to 1V scale (I also had tried 500mV, and it didn't seem to make a difference) and 10s time axis spacing (maximum integration time, since we're looking for a DC effect. Is this what we want?)
- The impedance for both channels is 1MOhm.
- I ran the script to start the loss measurement on the Y arm.
- python2 armloss_dcrefl_asdcpd_scope.py 192.168.113.24 1 2 5 YARM
- I'm reading ~15 (au?) for the MC channel and ~5% of that out the AS, which seems to make sense to me and looked to be about what Yuki the ratios when I checked the log files. However, I'm a bit confused by the normalization, because the maximum output of the MC PD is 10V, and indeed the scope's display is reading under 10V.
I've left the script running. |