Message ID: 14043     Entry time: Sat Jul 7 19:50:38 2018
Author: Annalisa 
Type: Configuration 
Category: Thermal Compensation 
Subject: Study about the Thermal projection setup and its effect on the cavity 

I made some simulation to study the change that the heater setup can induce on the Radius of Curvature of the ETM.

Heat pattern

First, I used a non-sequential ray tracing software (Zemax) to calculate the heat pattern. I made a CAD of the elliptical reflector and I put a radiative element inside it (similar to the rod-heater 30mm long, 3.8mm diameter that we ordered), placing it in such a way that the heater tip is as close as possible to the ellipse first focus. (figure 1)

Then, by putting a screen at the second focus of the ellipse (where we suppose to place the mirror HR surface), I could find the projected heat pattern, as shown in figure 2 and 3 (section). Notice that the scale is in INCH, even if the label says mm. As you can see, the heat pattern is pretty broad, but still enough to induce a RoC change. 

Mirror deformation

In order to compute the mirror deformation induced by this kind of pattern, I used this map produced with Zemax as absorption map in COMSOL. I considered ~1W total power absorbed by the mirror (just to have a unitary number).

The mirror temperature and deformation maps induced by this heat pattern are shown in figures 4 and 5. 

RoC change evaluation

Then I had to evaluate the RoC change. In particular, I did it by fitting the Radius of Curvature over a circle of radius:

r = w_{00} * \sqrt{n}

where w_{00} is the waist of tha Gaussian mode on the ETMY (5mm) and n is the mode order. This is a way to approximately know which is the Radius of Curvature as "seen" by each HOM, and is shown in figure 6 (the RoC of the cold mirror is set to be 57.37m). Of course, besides being very tiny, the difference in RoC strongly depends on the heat pattern.

Gouy phase variation

Considering this absorbed power, the cavity Gouy phase variation between hot and cold state is roughly 15kHz (I leave to the SURFs the details of the calculation).

Unanswered points

So the still unaswered questions are:

- which is the minimum variation we are able to resolve with our measurement

- how much heating power do we expect to be projected onto the mirror surface (I'll make another entry on that)

Attachment 1: reflector.png  30 kB  Uploaded Sat Jul 7 21:00:18 2018  | Hide | Hide all | Show all
reflector.png
Attachment 2: heat_pattern_-_f2.png  43 kB  Uploaded Sat Jul 7 21:01:05 2018  | Show | Hide all | Show all
Attachment 3: heat_pattern_-_f2_-_cross_section.png  35 kB  Uploaded Sat Jul 7 21:03:15 2018  | Show | Hide all | Show all
Attachment 4: ETMtemperature.png  74 kB  Uploaded Sat Jul 7 21:11:33 2018  | Hide | Hide all | Show all
ETMtemperature.png
Attachment 5: ETMdeformation.png  63 kB  Uploaded Sat Jul 7 21:11:39 2018  | Hide | Hide all | Show all
ETMdeformation.png
Attachment 6: RoC_variation.png  32 kB  Uploaded Sat Jul 7 22:02:01 2018  | Hide | Hide all | Show all
RoC_variation.png
ELOG V3.1.3-