40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sat Jun 23 13:08:41 2018, Jon, Update, AUX, First Coherent AUX Scan of PRC Using AM Sidebands IMG_2695.jpegIMG_2698.jpegIMG_2699.jpeg
    Reply  Sat Jun 23 20:54:35 2018, Koji, Update, AUX, First Coherent AUX Scan of PRC Using AM Sidebands 
    Reply  Tue Jun 26 10:06:39 2018, keerthana, Update, AUX, First Coherent AUX Scan of PRC Using AM Sidebands TRANS.pdfREFL.pdf
Message ID: 14011     Entry time: Sat Jun 23 20:54:35 2018     In reply to: 14010
Author: Koji 
Type: Update 
Category: AUX 
Subject: First Coherent AUX Scan of PRC Using AM Sidebands 

How much was the osc freq of the marconi? And then how much was the resulting freq offset between PSL and AUX?

Are we supposed to see two dips with the separation of an FSR? Or four dips (you have two sidebands)?

And the distance between the dips (28MHz-ish?) seems too large to be the FSR (22MHz-ish).
cf https://wiki-40m.ligo.caltech.edu/IFO_Modeling/RC_lengths

ELOG V3.1.3-