Entry  Mon Jun 11 23:22:53 2018, gautam, Update, CDS, EX wiring confusion 
    Reply  Wed Jun 13 23:23:44 2018, johannes, Update, CDS, EX wiring confusion c1auxex_channels.pdfXEND_slow_wiring.pdf
       Reply  Thu Jun 14 10:41:00 2018, gautam, Update, CDS, EX wiring confusion 
          Reply  Thu Jun 14 15:31:18 2018, johannes, Update, CDS, EX wiring confusion XEND_slow_wiring.pdfc1auxex_channels.pdf
Message ID: 13958     Entry time: Wed Jun 13 23:23:44 2018     In reply to: 13947     Reply to this: 13961
Author: johannes 
Type: Update 
Category: CDS 
Subject: EX wiring confusion 

It's true.

I went through the wiring of the c1auxex crate today to disentangle the pin assignments. The full detail can be found in attachment #1, #2 has less detail but is more eye candy. The red flagged channels are now marked for removal at the next opportunity. This will free up DAQ channels as follows:

TYPE Total Available now Available after
ADC 24 2 14
DAC 16 8 12
BIO sinking 16 7 7
BIO sourcing 8 8 8

This should be enough for temperature sensing, NPRO diagnostics, and even eventual remote PDH control with new servo boxes.

Attachment 1: c1auxex_channels.pdf  23 kB  Uploaded Thu Jun 14 00:25:03 2018  | Hide | Hide all
c1auxex_channels.pdf
Attachment 2: XEND_slow_wiring.pdf  212 kB  Uploaded Thu Jun 14 00:46:46 2018  | Hide | Hide all
XEND_slow_wiring.pdf