Message ID: 13938     Entry time: Mon Jun 11 11:45:13 2018     Reply to this: 13941
Author: keerthana  
Type: Update 
Category: elog 
Subject: Comparison of the analytical and finesse values of TMS and FSR. 
Quantity Analytical Value Finesse Value Percentage Error
Free Spectral range (FSR) 3.893408 MHz 3.8863685 MHz 0.180 %
Transverse Mode Spacing (TMS) 1.195503 MHz 1.1762885 MHz 1.607 %

The values obtained from both analytical and finesse solution is given in the above table along with the corresponding percentage errors.finesse1.pdf

The parameters used for this calculation are listed below.

Parameter Value
length of the cavity (L) 38.5 m
Wavelength of the laser beam (\lambda) 1064 nm
Radius of curvature of ITM (R1) \infty
Radius of curvature of ETM (R2) 58 m

The cavity scan data obtained from Finesse is also attached here.