Message ID: 13931     Entry time: Fri Jun 8 00:36:54 2018     In reply to: 13924
Author: gautam 
Type: Update 
Category: PSL 
Subject: observing the resonance signal corresponding to the injected frequency. 

It isn't clear to me in the drawing where the Agilent is during this measurement. Over 40m of cabling, the loss of signal can be a few dB, and considering we don't have a whole lot of signal in the first place, it may be better to send the stronger RF signal (i.e. Marconi pickoff) over the long cable rather than the weak beat signal from the Transmission photodiode. 