Tue Dec 26 17:24:24 2017, Steve, Update, General, power outage
|
Thu Dec 28 17:22:02 2017, gautam, Update, General, power outage - CDS recovery
|
Thu Jan 4 14:39:50 2018, gautam, Update, General, power outage - timing error
|
Fri Jan 5 21:54:28 2018, rana, Update, General, power outage - timing error
|
Fri Jan 5 22:19:53 2018, gautam, Update, General, power outage - timing error
|
Sat Jan 6 18:27:37 2018, gautam, Update, General, power outage - IFO recovery 
|
Mon Apr 16 22:09:53 2018, Kevin, Update, General, power outage - BLRM recovery 
|
Tue Apr 17 09:57:09 2018, Steve, Update, General, seismometer interfaces 
|
Wed Apr 18 20:33:19 2018, Kevin, Update, General, seismometer interfaces  
|
Fri Apr 20 23:36:28 2018, Kevin, Update, PEM, Seismometer BLRMs  
|
Tue Apr 24 21:19:08 2018, Kevin, Update, PEM, Seismometer BLRMs   
|
Wed Apr 25 17:44:39 2018, Arnold, Update, PEM, PEM Anti-Alias wiring
|
Thu Apr 26 09:35:49 2018, Kevin, Update, PEM, PEM Anti-Alias wiring 
|
Thu Apr 26 20:22:21 2018, Kevin, Update, PEM, ADC common mode rejection with new seismometer connections 
|
Wed May 16 21:02:22 2018, Kevin, Update, PEM, ADC common mode rejection with new seismometer connections
|
Thu Jun 14 15:24:32 2018, Steve, Update, PEM, ADC DAC In Line Test Boards are in
|
Tue Jan 2 15:43:35 2018, Steve, Update, VAC, pumpdown after power outage
|
Wed Feb 7 15:50:42 2018, Steve, Update, VAC, IFO pressure monitoring
|
|
Message ID: 13755
Entry time: Mon Apr 16 22:09:53 2018
In reply to: 13510
Reply to this: 13756
|
Author: |
Kevin |
Type: |
Update |
Category: |
General |
Subject: |
power outage - BLRM recovery |
|
|
I've been looking into recovering the seismic BLRMs for the BS Trillium seismometer. It looks like the problem is probably in the anti-aliasing board. There's some heavy stuff sitting on top of it in the rack, so I'll take a look at it later when someone can give me a hand getting it out.
In detail, after verifying that there are signals coming directly out of the seismometer, I tried to inject a signal into the AA board and see it appear in one of the seismometer channels.
- I looked specifically at C1:PEM-SEIS_BS_Z_IN1 (Ch9), C1:PEM-SEIS_BS_X_IN1 (Ch7), and C1:PEM-ACC_MC2_Y_IN1 (Ch27). All of these channels have between 2000--3000 cts.
- I tried injecting a 200 mVpp signal at 1.7862 Hz into each of these channels, but the the output did not change.
- All channels have 0 cts when the power to the AA board is off.
- I then tried to inject the same signal into the AA board and see it at the output. The setup is shown in the first attachment. The second BNC coming out of the function generator is going to one of the AA board inputs; the 32 pin cable is coming directly from the output. All channels give 4.6 V when when the board is powered on regardless of wheter any signal is being injected.
- To verify that the AA board is likely the culprit, I also injected the same signals directly into the ADC. The setup is shown in the second attachment. The 32 pin cable is going directly to the ADC. When injecting the same signals into the appropriate channels the above channels show between 200--300 cts, and 0 cts when no signal is injected.
|
|
|
|
|