40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Mon Mar 19 23:44:00 2018, gautam, Update, IOO, IMC loop checkup 
|
Tue Mar 20 16:53:03 2018, gautam, Update, IOO, IMC loop checkup
|
Tue Mar 20 21:08:03 2018, gautam, Update, IOO, IMC loop checkup
|
Tue Mar 20 22:44:45 2018, gautam, Update, IOO, IMC loop checkup
|
Wed Mar 21 15:52:45 2018, gautam, Update, IOO, MC error point calibration  
|
Wed Mar 21 17:31:21 2018, gautam, Update, IOO, MC error point calibration
|
Wed Mar 21 21:13:44 2018, gautam, Update, IOO, IMC noise budget
|
Tue Mar 27 01:39:44 2018, gautam, Update, IOO, PSL noise eater was off
|
Tue Mar 27 19:32:03 2018, arijit, Update, IOO, PSL noise eater was off 
|
Tue Mar 27 23:37:35 2018, gautam, Update, IOO, MC REFL PD removed
|
Wed Mar 28 21:31:39 2018, gautam, Update, IOO, MC REFL PD removed
|
Wed Mar 28 21:47:37 2018, arijit, Update, IOO, MCREFL_PD Optical response measurement 
|
Thu Mar 29 17:57:36 2018, arijit, Update, IOO, MCREFL_PD Optical response measurement
|
Fri Mar 30 06:14:31 2018, rana, Update, IOO, MCREFL_PD Optical response measurement
|
Fri Mar 30 22:37:36 2018, Kevin, Update, IOO, MCREFL_PD Optical response measurement
|
Thu Apr 5 04:36:56 2018, Kevin, Update, IOO, Coil driver noise  
|
Thu Apr 5 12:13:18 2018, Koji, Update, IOO, Coil driver noise
|
Fri Apr 6 21:39:09 2018, gautam, Update, IOO, MC2 suspension health checkup
|
Mon Apr 9 18:46:03 2018, gautam, Update, IOO, Further debugging 
|
Tue Apr 10 14:28:44 2018, gautam, Update, IOO, Further debugging
|
Wed Apr 11 01:34:31 2018, gautam, Update, IOO, Activities today
|
Thu Apr 12 18:12:49 2018, gautam, Update, ALS, NPRO channels hijacked  
|
Fri Apr 13 11:02:41 2018, gautam, Update, ALS, EX fiber polarization drift
|
Fri Apr 13 17:56:26 2018, gautam, Update, ALS, Fibers switched out
|
Sat Apr 14 14:42:09 2018, gautam, Update, ALS, Fibers switched out
|
Tue Apr 17 14:08:29 2018, gautam, Update, ALS, Fibers switched out
|
Fri Apr 20 00:26:34 2018, gautam, Update, ALS, Fibers switched out
|
Sat Apr 21 20:25:12 2018, gautam, Update, ALS, PSL fiber pickoff status
|
Thu Apr 19 00:03:51 2018, gautam, Update, IOO, More IMC NBing 
|
Thu Apr 19 17:15:35 2018, gautam, Update, IOO, More IMC NBing 
|
Sun May 20 23:43:50 2018, gautam, Update, IOO, Coil driver noise re-measurement    
|
Tue May 22 17:26:25 2018, gautam, Update, IOO, MC1 Coil Driver pulled out
|
Tue May 22 23:28:01 2018, gautam, Update, IOO, MC1 Coil Driver pulled out
|
Wed May 23 17:58:48 2018, gautam, Update, IOO, MC1 Coil Driver pulled out 
|
Wed May 30 10:17:46 2018, gautam, Update, IOO, MC1 Coil Driver pulled out 
|
Fri Apr 6 22:23:53 2018, Kevin, Update, IOO, Coil driver noise
|
|
Message ID: 13754
Entry time: Sat Apr 14 14:42:09 2018
In reply to: 13753
Reply to this: 13757
|
Author: |
gautam |
Type: |
Update |
Category: |
ALS |
Subject: |
Fibers switched out |
|
|
It looks like the drift in polarization content in the PSL pickoff is actually much higher than that in the EX pickoff. Note that to prevent the P-pol diode from saturating, I put an ND filter in front of the PD, so the Y axis actually has to be multiplied by 10 to compare power in S and P polarizations. If this drift is because of the input (linear) polarization being poorly matched to one of the fiber's special axes, then we can improve the situation relatively easily. But if the polarization drift is happening as a result of time-varying stress (due to temp. fluctuations, acoustics etc) on the (PM) fiber from the PSL fiber coupler to the BeatMouth, then I think this is a much more challenging problem to solve.
I'll attempt to quantify the contribution (in Hz/rtHz) of beat amplitude RIN to phase tracker output noise, which will tell us how much of a problem this really is and in which frequency bands. In particular, I'm interested in seeing if the excess noise around 100 Hz is because of beat amplitude fluctuations. But on the evidence thus far, I've seen the beat amplitude drift by ~15 dB (over long timescales) on the control room network analyzer, and this drift seems to be dominated by PSL light amplitude fluctuations. |
|
|