40m Log |
Not logged in |
 |
|
Mon Mar 19 23:44:00 2018, gautam, Update, IOO, IMC loop checkup 
|
Tue Mar 20 16:53:03 2018, gautam, Update, IOO, IMC loop checkup
|
Tue Mar 20 21:08:03 2018, gautam, Update, IOO, IMC loop checkup
|
Tue Mar 20 22:44:45 2018, gautam, Update, IOO, IMC loop checkup
|
Wed Mar 21 15:52:45 2018, gautam, Update, IOO, MC error point calibration  
|
Wed Mar 21 17:31:21 2018, gautam, Update, IOO, MC error point calibration
|
Wed Mar 21 21:13:44 2018, gautam, Update, IOO, IMC noise budget
|
Tue Mar 27 01:39:44 2018, gautam, Update, IOO, PSL noise eater was off
|
Tue Mar 27 19:32:03 2018, arijit, Update, IOO, PSL noise eater was off 
|
Tue Mar 27 23:37:35 2018, gautam, Update, IOO, MC REFL PD removed
|
Wed Mar 28 21:31:39 2018, gautam, Update, IOO, MC REFL PD removed
|
Wed Mar 28 21:47:37 2018, arijit, Update, IOO, MCREFL_PD Optical response measurement 
|
Thu Mar 29 17:57:36 2018, arijit, Update, IOO, MCREFL_PD Optical response measurement
|
Fri Mar 30 06:14:31 2018, rana, Update, IOO, MCREFL_PD Optical response measurement
|
Fri Mar 30 22:37:36 2018, Kevin, Update, IOO, MCREFL_PD Optical response measurement
|
Thu Apr 5 04:36:56 2018, Kevin, Update, IOO, Coil driver noise  
|
Thu Apr 5 12:13:18 2018, Koji, Update, IOO, Coil driver noise
|
Fri Apr 6 21:39:09 2018, gautam, Update, IOO, MC2 suspension health checkup
|
Mon Apr 9 18:46:03 2018, gautam, Update, IOO, Further debugging 
|
Tue Apr 10 14:28:44 2018, gautam, Update, IOO, Further debugging
|
Wed Apr 11 01:34:31 2018, gautam, Update, IOO, Activities today
|
Thu Apr 12 18:12:49 2018, gautam, Update, ALS, NPRO channels hijacked  
|
Fri Apr 13 11:02:41 2018, gautam, Update, ALS, EX fiber polarization drift
|
Fri Apr 13 17:56:26 2018, gautam, Update, ALS, Fibers switched out
|
Sat Apr 14 14:42:09 2018, gautam, Update, ALS, Fibers switched out
|
Tue Apr 17 14:08:29 2018, gautam, Update, ALS, Fibers switched out
|
Fri Apr 20 00:26:34 2018, gautam, Update, ALS, Fibers switched out
|
Sat Apr 21 20:25:12 2018, gautam, Update, ALS, PSL fiber pickoff status
|
Thu Apr 19 00:03:51 2018, gautam, Update, IOO, More IMC NBing 
|
Thu Apr 19 17:15:35 2018, gautam, Update, IOO, More IMC NBing 
|
Sun May 20 23:43:50 2018, gautam, Update, IOO, Coil driver noise re-measurement    
|
Tue May 22 17:26:25 2018, gautam, Update, IOO, MC1 Coil Driver pulled out
|
Tue May 22 23:28:01 2018, gautam, Update, IOO, MC1 Coil Driver pulled out
|
Wed May 23 17:58:48 2018, gautam, Update, IOO, MC1 Coil Driver pulled out 
|
Wed May 30 10:17:46 2018, gautam, Update, IOO, MC1 Coil Driver pulled out 
|
Fri Apr 6 22:23:53 2018, Kevin, Update, IOO, Coil driver noise
|
|
Message ID: 13738
Entry time: Fri Apr 6 22:23:53 2018
In reply to: 13730
|
Author: |
Kevin |
Type: |
Update |
Category: |
IOO |
Subject: |
Coil driver noise |
|
|
Quote: |
Why is MC2 LR so different from the others???
|
The previous measurements were made from the coil driver output monitors. To investigate why the MC2 LR coil has less noise than the other coils, I also measured the noise at the output to the coils.
The circuit diagram for the coil driver board is given in D010001 and a picture of the rack is on the 40m wiki here. The coil driver boards are in the upper left quadrant of the rack. The input to the board is the column of LEMOs which are the "Coil Test In" inputs on the schematic. The output monitors are the row of LEMOs to the right of the input LEMOs and are the "FP Coil Volt Mon" outputs on the schematic. The output to the coils "Coil Out" in the schematic are carried through a DB15 connector.
The attachment shows the voltage noise for the MC2 LR coil as well as the UL coil which is similar to all of the other coils measured in the previous measurement. While the LR coil is less noisy than the UL coil as measured at the output monitor, they have the same noise spectrum as measured at the output to the coils themselves. So there must be something wrong with the buffer circuit for the MC2 LR voltage monitor, but the output to the coils themselves is the same as for the other coils. |
|
|