40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Oct 9 01:29:26 2017, gautam, Update, LSC, DRMI Nosie Budget v3.0 7x
    Reply  Mon Oct 9 11:55:01 2017, Koji, Update, LSC, DRMI Nosie Budget v3.0 
       Reply  Mon Oct 9 22:18:34 2017, gautam, Update, LSC, AS55Q Dark Noise AS55Q_dark.pdf
          Reply  Tue Oct 10 22:04:06 2017, rana, Update, LSC, AS55Q Dark Noise 
          Reply  Wed Oct 11 14:42:03 2017, gautam, Update, LSC, AS55Q Dark Noise AS55Q_darkNoises.pdf
             Reply  Wed Oct 11 19:31:32 2017, gautam, Update, LSC, AS55Q Dark Noise AS55Q_darkNoises.pdf
                Reply  Thu Oct 12 01:50:11 2017, gautam, Update, LSC, AS55Q Dark Noise AS55_dark.png
                   Reply  Thu Oct 12 12:17:28 2017, gautam, Update, LSC, AS55Q Dark Noise D040179_AS55_40m.pdf
                      Reply  Fri Oct 13 12:26:12 2017, gautam, Update, LSC, AS55Q Dark Noise AS55_DemodNoises.pdf
                         Reply  Mon Oct 16 16:01:04 2017, gautam, Update, LSC, AS55Q Dark Noise DemodBoardwOscope.pdfDemodBoardwOscope_ASD.pdf
                            Reply  Tue Oct 17 19:31:53 2017, gautam, Update, LSC, AS55Q Dark Noise D040179_TFs.pdfAS55_DemodNoises.pdf
    Reply  Tue Nov 7 17:45:05 2017, gautam, Update, LSC, DRMI Nosie Budget v3.1 SRCL_MICH_whitenoise_tf.pdfPRCL_MICH_whitenoise_tf.pdfMICH_aux.pdfC1NB_disp_40m_MICH_NB_2017-10-08.pdf
       Reply  Tue Nov 7 22:56:21 2017, gautam, Update, LSC, DRMI locking recovered DRMI_7Nov20178.png
          Reply  Wed Nov 8 00:28:16 2017, gautam, Update, LSC, Laser intensity coupling measurement attempt PSL_RIN.pdf
       Reply  Wed Nov 8 09:37:45 2017, rana, Update, LSC, DRMI Nosie Budget v3.1 
          Reply  Wed Nov 8 09:59:12 2017, gautam, Update, LSC, DRMI Nosie Budget v3.1 
             Reply  Wed Nov 15 01:37:07 2017, gautam, Update, LSC, DRMI low freq. nosie improved C1NB_disp_40m_MICH_NB_2017-11-15.pdfOL_A2L_couplings.pdf
                Reply  Thu Nov 16 00:53:26 2017, gautam, Update, LSC, DRMI noise sub-budgets C1NB_a2l_40m_MICH_NB_2017-11-15.pdfC1NB_aux_40m_MICH_NB_2017-11-15.pdf
Message ID: 13374     Entry time: Wed Oct 11 19:31:32 2017     In reply to: 13372     Reply to this: 13376
Author: gautam 
Type: Update 
Category: LSC 
Subject: AS55Q Dark Noise 

I tried replacing the AD797s on the daughter board with OP27s, and saw no significant improvement in the electronics noise of the demod board. Note that according to LISO, in this configuration, the voltage noise of the Op27 is expected to dominate the total noise of the daughter board. Measurement condition was that the RFPD input was terminated, but the LO input was still being driven (SR785 input range is -50dBVpk for all traces, and the input ranging was set to "UpOnly"). Need to do a more systematic investigation to figure out where this excess noise is coming from. I will upload photos of the board later.

Quote:

This supports the hypothesis that something is wonky on the daughter board, because the purple trace should only be the quad sum of the orange and green traces. I will pull it out and have a look.

 

Attachment 1: AS55Q_darkNoises.pdf  167 kB  Uploaded Wed Oct 11 20:34:46 2017  | Show | Show all