40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Oct 9 01:29:26 2017, gautam, Update, LSC, DRMI Nosie Budget v3.0 7x
    Reply  Mon Oct 9 11:55:01 2017, Koji, Update, LSC, DRMI Nosie Budget v3.0 
       Reply  Mon Oct 9 22:18:34 2017, gautam, Update, LSC, AS55Q Dark Noise AS55Q_dark.pdf
          Reply  Tue Oct 10 22:04:06 2017, rana, Update, LSC, AS55Q Dark Noise 
          Reply  Wed Oct 11 14:42:03 2017, gautam, Update, LSC, AS55Q Dark Noise AS55Q_darkNoises.pdf
             Reply  Wed Oct 11 19:31:32 2017, gautam, Update, LSC, AS55Q Dark Noise AS55Q_darkNoises.pdf
                Reply  Thu Oct 12 01:50:11 2017, gautam, Update, LSC, AS55Q Dark Noise AS55_dark.png
                   Reply  Thu Oct 12 12:17:28 2017, gautam, Update, LSC, AS55Q Dark Noise D040179_AS55_40m.pdf
                      Reply  Fri Oct 13 12:26:12 2017, gautam, Update, LSC, AS55Q Dark Noise AS55_DemodNoises.pdf
                         Reply  Mon Oct 16 16:01:04 2017, gautam, Update, LSC, AS55Q Dark Noise DemodBoardwOscope.pdfDemodBoardwOscope_ASD.pdf
                            Reply  Tue Oct 17 19:31:53 2017, gautam, Update, LSC, AS55Q Dark Noise D040179_TFs.pdfAS55_DemodNoises.pdf
    Reply  Tue Nov 7 17:45:05 2017, gautam, Update, LSC, DRMI Nosie Budget v3.1 SRCL_MICH_whitenoise_tf.pdfPRCL_MICH_whitenoise_tf.pdfMICH_aux.pdfC1NB_disp_40m_MICH_NB_2017-10-08.pdf
       Reply  Tue Nov 7 22:56:21 2017, gautam, Update, LSC, DRMI locking recovered DRMI_7Nov20178.png
          Reply  Wed Nov 8 00:28:16 2017, gautam, Update, LSC, Laser intensity coupling measurement attempt PSL_RIN.pdf
       Reply  Wed Nov 8 09:37:45 2017, rana, Update, LSC, DRMI Nosie Budget v3.1 
          Reply  Wed Nov 8 09:59:12 2017, gautam, Update, LSC, DRMI Nosie Budget v3.1 
             Reply  Wed Nov 15 01:37:07 2017, gautam, Update, LSC, DRMI low freq. nosie improved C1NB_disp_40m_MICH_NB_2017-11-15.pdfOL_A2L_couplings.pdf
                Reply  Thu Nov 16 00:53:26 2017, gautam, Update, LSC, DRMI noise sub-budgets C1NB_a2l_40m_MICH_NB_2017-11-15.pdfC1NB_aux_40m_MICH_NB_2017-11-15.pdf
Message ID: 13368     Entry time: Mon Oct 9 11:55:01 2017     In reply to: 13367     Reply to this: 13369
Author: Koji 
Type: Update 
Category: LSC 
Subject: DRMI Nosie Budget v3.0 

My last characterization of the AS55 PD was on Feb 2013. ELOG 8100

There I said the dark noise at the PD output was 16nV/rtHz. I don't have the measurement of the Voltage noise at the output of the demod board.

Note that the PD can only be limited by shot noise when the DC current is larger then 4mA.