40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Sep 11 23:31:50 2017, johannes, Update, Cameras, post-vent camera capture comparison 13x
    Reply  Tue Sep 26 22:11:08 2017, johannes, Update, Cameras, post-vent camera capture comparison lens_distance.pdfImage__2017-09-26__18-54-29.jpgIMG_20170926_190157222_HDR.jpgIMG_20170926_185649325.jpg
       Reply  Mon Oct 2 12:44:45 2017, johannes, Update, Cameras, Basler 120gm calibration calib_20170930_152.pdfcalib_20170930_153.pdf
          Reply  Mon Oct 2 23:16:05 2017, gautam, HowTo, Cameras, CCD calibration 
             Reply  Tue Oct 3 01:58:32 2017, johannes, HowTo, Cameras, CCD calibration gige_calibration.pdf
                Reply  Thu Oct 12 01:03:49 2017, johannes, HowTo, Cameras, ETMX GigE side view IMG_20171011_164549698.jpgImage__2017-10-11__16-52-01.pngGigE_lens_position_helper.py.zip
                   Reply  Thu Oct 12 07:56:33 2017, Steve, HowTo, Cameras, ETMX GigE side view at 50 deg of IR scattering Image__2017-10-11__15-29-52_15k400g.pngImage__2017-10-12__15-50-18wipedRefocud2.pngcamCan16cm.jpg
                      Reply  Wed Oct 18 11:37:58 2017, johannes, HowTo, Cameras, ETMX GigE side view at 50 deg 
                Reply  Wed Oct 18 15:26:58 2017, johannes, HowTo, Cameras, Revision: CCD calibration calib_20170930_152.pdfcalib_20170930_153.pdf
Message ID: 13334     Entry time: Tue Sep 26 22:11:08 2017     In reply to: 13310     Reply to this: 13348
Author: johannes 
Type: Update 
Category: Cameras 
Subject: post-vent camera capture comparison 

I configured the remaining GigE-Camera to work on the 40m network. We currently have 3 operational Basler cameras:

The 120gm's have been assigned the IPs 192.168.113.152  (was already configured) and 192.168.113.153 (freshly configured) and have been labeled accordingly. Note that it was not necessary to connect the out-of-the-box camera directly to a dedicated ethernet adapter whose IP was set manually to 169.254.0.XXX as pointed out in earlier posts - a few seconds after connecting the camera to the control room switch (with PoE adapter to power it) the camera showed up in the configuration software tool which is launched via

/opt/rtcds/caltech/c1/scripts/GigE/pylon5/bin/./IpConfigurator

and can be assigned a corrected, static IP.

We have a plethora of 2" tubes for the lens assembly, but not a great variety of focal lengths for 2" lenses. Present with the camera gear were two f=250 mm and one f=150 mm 2" lenses with a NIR broadband AR coating

To determine the lens positions relativ to the sensor I assumed that the camera we're setting up looks at its test mass from a distance of 1m. Using the two available focal lengths we can look for solutions which have reasonable lens separations <~10cm and suitable magnification. We primarily want to image the central mirror area onto a 1/4" sized sensor, which can be achieved with a magnification of ~1/8.

I chose a lens separation of 6cm, which gives a theoretical magnification of -.12 and a sensor-lens 2 distance of 7.95 cm. I placed the lenses accordingly in the tubes and checked the focusing with Gautam's help:

       

It's pretty close to what we would expect. We will do the calibration using the auxiliary laser on the PSL table. For this I temporarily routed a fiber from the PSL enclosure to the SP table. Since the main cable hole is sort of cramped it's going in through a gap near the ceiling instead.  

 

Attachment 1: lens_distance.pdf  14 kB  | Hide | Hide all
lens_distance.pdf