Mon Mar 6 12:34:30 2017, johannes, Summary, ASS, ASS light injection scenarios
|
Mon Aug 21 20:11:25 2017, johannes, Summary, General, Loss measurements plan
|
Mon Aug 21 21:26:41 2017, gautam, Summary, General, Loss measurements plan
|
Tue Aug 22 16:56:54 2017, johannes, Summary, General, AS laser existing components inventory
|
Mon Aug 28 23:22:56 2017, johannes, Update, PSL, PSL table auxiliary NPRO   
|
Tue Aug 29 20:04:09 2017, rana, Update, PSL, PSL table auxiliary NPRO
|
Tue Aug 29 21:36:59 2017, johannes, Update, PSL, PSL table auxiliary NPRO 
|
Tue Sep 5 23:13:44 2017, johannes, Update, PSL, PSL table auxiliary NPRO 
|
Thu Sep 7 23:09:00 2017, johannes, Update, PSL, PSL table auxiliary NPRO 
|
Mon Sep 11 12:40:32 2017, johannes, Update, PSL, PSL table auxiliary NPRO 
|
|
Message ID: 13306
Entry time: Mon Sep 11 12:40:32 2017
In reply to: 13301
|
Author: |
johannes |
Type: |
Update |
Category: |
PSL |
Subject: |
PSL table auxiliary NPRO |
|
|
I changed the PSL table auxiliary laser setup to the 200 MHz AOM and put the light back in the fiber. Coupling efficiency is again ~50%, giving us up to about 75 mW of auxiliary laser light on the AS table. The 90% to 10% fall time of the light power out of the fiber when switched off is 16.5 ns with this AOM on the PDA10A, which will be sufficient for the ringdown measurements.


|